Zimmer Manufacturing B.V. recalls Product 14 consists of all product under product code HSB , and same usage: Item no: 225600230 ASIA ONE-PIECE LAG S

Recalled March 23, 2016 · FDA recall Z-1112-2016 · Zimmer Manufacturing B.V

Hazard
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile. · FDA Class II
Units
0
Sold at
US Nationwide Distribution

Description

As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

Product
Product 14 consists of all product under product code HSB , and same usage: Item no: 225600230 ASIA ONE-PIECE LAG SCREW 225600232 ASIA ONE-PIECE LAG SCREW 225600235 ASIA ONE-PIECE LAG SCREW 225600237 ASIA ONE-PIECE LAG SCREW 225600240 ASIA ONE-PIECE LAG SCREW 225600242 ASIA ONE-PIECE LAG SCREW 225600245 ASIA ONE-PIECE LAG SCREW 225600227 ASIA ONE-PIECE LAG SCREW 70MM Product Usage: An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Manufacturer(s)
Zimmer Manufacturing B.V.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov