RecallWatchUSA

Home / Recalls / Medical Devices

Zimmer Manufacturing B.V. recalls Product 50 consists of all product under product code: JWH and same usage: Item no: 595002701 NEXGEN MIS TIBIA PLATE

Recalled March 23, 2016 · FDA recall Z-1148-2016 · Zimmer Manufacturing B.V

Hazard
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile. · FDA Class II
Units
5567
Sold at
US Nationwide Distribution

Description

As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

Product
Product 50 consists of all product under product code: JWH and same usage: Item no: 595002701 NEXGEN MIS TIBIA PLATE PR 595002702 NEXGEN MIS TIBIA PLATE PR 595003701 NEXGEN MIS TIBIA PLATE PR 595003702 NEXGEN MIS TIBIA PLATE PR 595003712 NEXGEN MIS TIBIA PLATE PR 595004701 NEXGEN MIS TIBIA PLATE PR 595004702 NEXGEN MIS TIBIA PLATE PR 595005701 NEXGEN MIS TIBIA PLATE PR 595005702 NEXGEN MIS TIBIA PLATE PR for use in total knee arthroplasty
Manufacturer(s)
Zimmer Manufacturing B.V.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Zimmer Manufacturing B.V. recalls Product 50 consists of all product under product code: JWH and same usage: Item no: 595002701 NEXGEN MIS TIBIA PLATE | RecallWatchUSA