Zyno Medical LLC recalls Zyno Medical Administration Set -105"ADMIN SET 20DP W/FILTER 2YSITES CLMP- For Use with Zyno Medical Infusion pumps or G

Recalled July 31, 2019 · FDA recall Z-2059-2019 · Zyno Medical

Hazard
Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified · FDA Class II
Units
360 cases (50/case)=18,000
Sold at
MA

Description

Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified

Product
Zyno Medical Administration Set -105"ADMIN SET 20DP W/FILTER 2YSITES CLMP- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code: B2-70072-F
Manufacturer(s)
Zyno Medical LLC