RecallWatchUSA

Home / Recalls / Medical Devices

Zyno Medical LLC recalls Zyno Medical Z800, Z800F, Z800W, and Z800WF Infusion Pumps

Recalled July 10, 2024 · FDA recall Z-2263-2024 · Zyno Medical

Hazard
Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having insufficient or no charge available to the pump when it is unplugged from the AC mains power. · FDA Class II
Units
88 units
Sold at
US Nationwide distribution.

Description

Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having insufficient or no charge available to the pump when it is unplugged from the AC mains power.

Product
Zyno Medical Z800, Z800F, Z800W, and Z800WF Infusion Pumps
Manufacturer(s)
Zyno Medical LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov