March 19, 2014Medical DevicesFDA
Accumetrics Inc recalls VerifyNow IIb/IIIa Test, Catalog No. 85310, 10-Test Kit and Catalog No. 85011, 25-Test kit. Product Usage: The Veri
Hazard: Accumetrics is recalling the VerifyNow IIb/IIIa test because it may result in the reporting of an erroneous low platelet aggregation unit (PAU) result. An erroneous low PAU result may cause a low percent inhibition calculation or low baseline PAU.
December 18, 2013Medical DevicesFDA
Accumetrics Inc recalls VerifyNow IIb/IIIa Test, 10-Test Kit, Catalog No. 85310, 10 tests/box; 690 total tests. The VerifyNow IIb/IlIa Tes
Hazard: Accumetrics is recalling the VerifyNow IIb/IIIa 10-Test because it contains incorrect information for the amount of time in which to perform the test following collection of the patient sample. The correct instruction is to perform the test within 15 minutes of collecting the patient sample.
June 27, 2012Medical DevicesFDA
Accumetrics Inc recalls VerifyNow System, part #85005-6H Product Usage: The VerifyNow P2Y12 assay device is a whole blood assay used in the
Hazard: The recall was initiated by Accumetrics because of a field correction to the VerifyNow System; specifically the results reported when running the VerifyNow P2Y12 Test. Currently the VerifyNow P2Y12 Test reports three values: PRU, BASE and % Inhibition. The change implemented is the elimination of the BASE and % Inhibition results. There will be no change to the reporting of the PRU result whic
June 27, 2012Medical DevicesFDA
Accumetrics Inc recalls VerifyNow P2Y12 Assay, Part Number: 85064, All reagent lots. Product Usage: The VerifyNow P2Y12 assay device is a wh
Hazard: The recall was initiated by Accumetrics because of a field correction to the VerifyNow System; specifically the results reported when running the VerifyNow P2Y12 Test. Currently the VerifyNow P2Y12 Test reports three values: PRU, BASE and % Inhibition. The change implemented is the elimination of the BASE and % Inhibition results. There will be no change to the reporting of the PRU result whic