Accumetrics Inc recalls VerifyNow P2Y12 Assay, Part Number: 85064, All reagent lots. Product Usage: The VerifyNow P2Y12 assay device is a wh
Recalled June 27, 2012 · FDA recall Z-1853-2012 · Accumetrics
Hazard
The recall was initiated by Accumetrics because of a field correction to the VerifyNow System; specifically the results reported when running the VerifyNow P2Y12 Test. Currently the VerifyNow P2Y12 Test reports three values: PRU, BASE and % Inhibition. The change implemented is the elimination of the BASE and % Inhibition results. There will be no change to the reporting of the PRU result whic · FDA Class II
Units
26,075 units
Sold at
US Nationwide Distribution
Description
The recall was initiated by Accumetrics because of a field correction to the VerifyNow System; specifically the results reported when running the VerifyNow P2Y12 Test. Currently the VerifyNow P2Y12 Test reports three values: PRU, BASE and % Inhibition. The change implemented is the elimination of the BASE and % Inhibition results. There will be no change to the reporting of the PRU result whic
Product
VerifyNow P2Y12 Assay, Part Number: 85064, All reagent lots. Product Usage: The VerifyNow P2Y12 assay device is a whole blood assay used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade. The VerifyNow Systems intended for use with human whole blood and verify now test devices. The system should be operated by healthcare professionals trained on use of the system and in accordance with institution and policies and procedures. Accumetrics respresentatives will assist your institution in the installation of the system and operations of the operator.
Manufacturer(s)
Accumetrics Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →