Alere San Diego Recalls

49 recalls on record · July 18, 2012 to May 8, 2024 · Medical Devices

Recalls by year

12
13
14
15
16
18
19
22
24

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

Also filed under

The recall feeds record what looks like the same company under more than one name:

March 26, 2014Medical DevicesFDA
Alere San Diego, Inc. recalls Product Brand Name(s): Alere Triage¿ TOX+MTD Drug Screen, Catalog Number 94400, Lots W56013B and W56035B. 25 individu
Hazard: Alere San Diego, Inc. is initiating a recall of the Alere Triage¿ TOX Drug Screen, Catalog Number 94400, Lots W56013B and W56035B because Alere's investigation has shown that these lots have a frequency of false negative results for PCP (Phencyclidine) that is higher than the frequency shown in the package insert. If you have questions about previously reported results.
January 15, 2014Medical DevicesFDA
Alere San Diego, Inc. recalls Product Brand Name(s): Innovacon Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device, OrALert Ora
Hazard: Alere initiated this recall of certain lots of the Oral Fluid Drug Screen Device because the sponge may become dislodged from the Saliva Collector handle prior to or during the oral fluid sample collection process. If the sponge becomes dislodged during the specimen collection process, the donor may inadvertently swallow the sponge which could present a choking risk.
December 4, 2013Medical DevicesFDA
Alere San Diego, Inc. recalls Product Brand Name(s): Alere Cholestech LDX¿ Lipid Profile"GLU Test Cassette, Model 10-991. Product Generic Name: Alere
Hazard: Alere initiated a recall of Cholestech LDX Lipid Profile-GLU test cassettes, Catalog Number 10-991, Lot # 320167, becuase an incorrect lot number was printed on the outer 10 pack kit box label. The incorrect Lot # number printed was 321067.
October 30, 2013Medical DevicesFDA
Alere San Diego, Inc. recalls Alere Triage¿ TOX Drug Screen Control 1 Product Generic Name: Quality Control Material Description of the product:
Hazard: Alere San Diego, Inc. initiated a voluntary recall of Alere Triage TOX Drug Screen Control 1 (PN 94413, Lot C2681A) because the CODE CHIP that was provided with this lot causes the control to fail when run in the QC Sample Mode using Alere Triage TOX Drug Screeen products. The Alere Triage TOX Drug Screen test devices and patient results are not affected.
June 5, 2013Medical DevicesFDA
Alere San Diego, Inc. recalls Alere" hCG Combo Cassette (20/10 mIU/mL) Kit. Product Usage: The AlereTM hCG Combo (20/10 mIU/mL) test is a rapid
Hazard: The firm initiated recall of Alere" hCG Combo Cassette (20/10 mIU/mL) kit PN 92215, lots hCG1110133 and hCG1110135, because an investigation has indicated a portion of these Alere" hCG Combo Cassette (20/10 mIU/mL) PN 92215 kit lots were incorrectly packaged with the Alere" hCG Cassette (20 mIU/mL) PN 92210 Product Insert. The Alere" hCG Cassette (20 mIU/mL) Product Insert contains information for
May 29, 2013Medical DevicesFDA
Alere San Diego, Inc. recalls Cardinal Health hCG Cassette Rapid Test, Part Number: B1077-22 (Mfg PN: 4582105026) The SP hCG Cassette Rapid Test is
Hazard: Alere San Diego Inc. is recalling the Cardinal Health hCG Cassette Rapid Test because some of them were incorrectly packaged into Cardinal Health hCG Combo Rapid test kit boxes. The kit box incorrectly indicates the product may be used for both urine and serum sample types.
May 1, 2013Medical DevicesFDA
Alere San Diego, Inc. recalls The Alere Cholestech LDX¿ high sensitivity C-Reactive Protein (hsCRP) Test Cassette, Model Number 12-807. Product
Hazard: The Alere Cholestech LDX¿ hsCRP cassette (PN 12-807) may have increased imprecision relative to the performance data in the package insert, which may cause differences in quantitative results for hsCRP compared to a reference method or repeat testing on the same patient sample.
April 24, 2013Medical DevicesFDA
Alere San Diego, Inc. recalls IMO-402 Mononucleosis Rapid Test Device (Whole Blood/Serum/Plasma). Catalog #IMO-402/94201 The MONO test (Whole Bloo
Hazard: Alere San Diego is recalling the Clearview Mononucleosis Cassette because the positive control included in the rapid test may produce false negative results.
April 24, 2013Medical DevicesFDA
Alere San Diego, Inc. recalls Alere Cholestech LDX ALT AST Test Cassette, Model #12-788. For the in vitro quantitative determination of alanine ami
Hazard: The recall was initiated because Alere San Diego has an update for the Alere Cholestech LDX ALT AST cassette, and it has the potential for humidity changes to impact results for certain analytes on the Alere Cholestech LDX System.
February 6, 2013Medical DevicesFDA
Alere San Diego, Inc. recalls Alere Cholestech LDX Calibration Verification, 4 vial set, Catalog Number: 88770 Alere Cholestech LDX Calibration Ver
Hazard: Out of range (increased) HDL Cholesterol results.
February 6, 2013Medical DevicesFDA
Alere San Diego, Inc. recalls Alere Cholestech LDX Multianalyte Control, 2 vial set, Catalog Number: 88769 Alere Cholestech LDX Multianalyte Contro
Hazard: Out of range (increased) HDL Cholesterol results.
October 17, 2012Medical DevicesFDA
Alere San Diego, Inc. recalls Triage TOX Drug Screen Panel, Model# 94400. Product Usage: The Triage TOX Drug Screen is a fluorescence immunoassa
Hazard: Biosite is recalling the Triage TOX Drug Screen kits where the box containing test devices because it was mislabeled.
October 3, 2012Medical DevicesFDA
Alere San Diego, Inc. recalls Alere Triage TOX Drug Screen 9 Panel, P/N: 94402. The Alere Triage TOX Drug Screen is a fluorescence immunoassay inte
Hazard: Alere San Diego is recalling Alere Triage TOX Drug Screen because there is a possibility that after installing the code chip provided with the affected lots: AMP, BAR, and PCP, the threshold ranges may not be consistently set correctly in all meters.
July 25, 2012Medical DevicesFDA
Alere San Diego, Inc. recalls Alere Triage Tox+MTD Drug Screen, PN 94400 Product Usage: The Alere Triage TOX Drug Screen is a fluorescence immunoa
Hazard: A recall was initiated because Alere San Diego has confirmed that the Triage TOX Drug Screen may have significantly decreased precision and accuracy than claimed. An increased frequency of false positive and/or false negative results has been observed for Triage TOX Drug Screen.
July 18, 2012Medical DevicesFDA
Alere San Diego, Inc. recalls Triage D-dimer PN 98100 Used as an aid in the assessment and evaluation of patients suspected of having disseminated
Hazard: Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
July 18, 2012Medical DevicesFDA
Alere San Diego, Inc. recalls Alere Triage BNP PN 98000XR Intended to be used as an aid in the diagnosis and assessment of severity of congestive h
Hazard: Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
July 18, 2012Medical DevicesFDA
Alere San Diego, Inc. recalls Alere Triage Profiler SOB Panel PN 97300 Used for the determination of creatine kinase MB, myoglobin, troponin I, B-t
Hazard: Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
July 18, 2012Medical DevicesFDA
Alere San Diego, Inc. recalls Alere Triage Cardiac Panel PN 97000HS Used for the quantitative determination of creatine kinase MD (CK--MB), myoglob
Hazard: Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
July 18, 2012Medical DevicesFDA
Alere San Diego, Inc. recalls Alere Triage CardioProfiler Panel PN 97100CP Used for the quantitative determination of creatine kinase MB, myoglobin
Hazard: Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.