Ascom (US) Recalls
2 recalls on record · May 18, 2016 to November 28, 2018 · Medical Devices
Recalls by year
16
18
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
November 28, 2018Medical DevicesFDA
Ascom (US) Inc. recalls Ascom Telligence Nurse Call System
Hazard: There is an issue that could cause the Telligence System Gateway II, NGGTWY2 H, to stop communicating to the hallway devices connected to it. When this occurs, the dome lights connected to the Gateway II Ports A and B will start flashing, indicating that they are no longer communicating to the Gateway II. If this occurs, device failures that WOULD otherwise be annunciated WILL NOT be indicated at staff consoles that are configured to display failures.
May 18, 2016Medical DevicesFDA
Ascom US, INC. recalls Ascom Mobile Monitoring Gateway (versions 4.1.1 and 4.2.0) and Ascom Unite Connect for Nurse Call (version 2.2.0).
Hazard: Due to a malfunction of software, the secondary module will not automatically take over messaging if the primary module fails after a software upgrade to one of the other affected versions. Messages from GE Carescape/Nurse Call will not forward to handset until primary module is restored.