Barco N.V Recalls

5 recalls on record · December 12, 2018 to July 2, 2025 · Medical Devices

Recalls by year

18
21
24
25

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

July 2, 2025Medical DevicesFDA
Barco N.V. recalls Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC -H/K9303300
Hazard: Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be interrupted or fall away, displays might switch to failover and back multiple times, devices/sources/sinks might not be discovered/available, might lead to delayed/prolonged procedures
July 17, 2024Medical DevicesFDA
Barco N.V. recalls MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display
Hazard: We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.
February 14, 2024Medical DevicesFDA
Barco N.V. recalls Barco MDSC-8527 NXF, Part Number K9352421
Hazard: Barco became aware of a problem with I2C cables in the manufacturing process of a limited batch of MDSC-85xx NXF products. The functionality of the built-in MNA decoder is blocked.
March 10, 2021Medical DevicesFDA
Barco N.V. recalls Barco MDNC-3421 DISPLAY, DIAGNOSTIC RADIOLOGY
Hazard: A number of MDNC-3421 displays were shipped while not being DICOM-compliant. These displays will not pass DICOM acceptance tests.
December 12, 2018Medical DevicesFDA
Barco N.V. recalls Optional MDRC-1119 Front Cover Attachment Kit for review display monitors. Product Usage: The product is an optiona
Hazard: If metal strips are removed and re-applied to another display, there is potential for the metal strips to be deformed resulting in a smaller contact surface with the front cover magnet strips. There is a risk that the protective front cover could come loose and may fall down.