RecallWatchUSA

Home / Recalls / Medical Devices

Barco N.V. recalls MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display

Recalled July 17, 2024 · FDA recall Z-2332-2024 · Barco N.V

Hazard
We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible. · FDA Class II
Units
88 units
Sold at
US: FL

Description

We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.

Product
MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display
Manufacturer(s)
Barco N.V.
Made in
Belgium

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Barco N.V. recalls MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display | RecallWatchUSA