3 recalls on record · October 31, 2012 to September 18, 2013 · Medical Devices
Recalls by year
12
13
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
September 18, 2013Medical DevicesFDA
Baxa Corporation recalls Baxa Vial Adapter, 14mm - Luer Lock, Order No. REF: 90, Quantity: 50 each, Single Use. Product Code H93890. Used as a
Hazard: Baxter Healthcare Corporation is voluntarily recalling one lot of Baxa Vial Adapters, 14mm - Luer Lock, due to incorrect expiry date.
July 24, 2013Medical DevicesFDA
Baxa Corporation recalls ExactaMix 1200 and 2400 Compounders. Used as an automated pumping system that compounds multiple sterile ingredients
Hazard: Baxter Healthcare is recalling certain ExactaMix automated compounding devices due to the complementary metal oxide semi-conductor battery failure resulting in resetting the internal clock of the device to midnight December 31, 2010.
October 31, 2012Medical DevicesFDA
Baxa Corporation recalls Repeater Pump, Class II device used for facilitating repeatable drug dosage distribution and reconstitution in hospital
Hazard: The firm is issuing a field correction to the operators manual included with the 099 Repeater Pump because the Technical Manual was distributed with the device as opposed to the Operator Manual and the instruction for use contradict each other with regards to multiple source container use.