Baxa Corporation recalls ExactaMix 1200 and 2400 Compounders. Used as an automated pumping system that compounds multiple sterile ingredients

Recalled July 24, 2013 · FDA recall Z-1752-2013 · Baxa

Hazard
Baxter Healthcare is recalling certain ExactaMix automated compounding devices due to the complementary metal oxide semi-conductor battery failure resulting in resetting the internal clock of the device to midnight December 31, 2010. · FDA Class II
Units
753 units
Sold at
Worldwide Distribution-USA (nationwide) including Puerto Rico, and the countries of Canada, Latin America, Europe/Middle East/Africa, and Taiwan.

Description

Baxter Healthcare is recalling certain ExactaMix automated compounding devices due to the complementary metal oxide semi-conductor battery failure resulting in resetting the internal clock of the device to midnight December 31, 2010.

Product
ExactaMix 1200 and 2400 Compounders. Used as an automated pumping system that compounds multiple sterile ingredients into a finished solution.
Manufacturer(s)
Baxa Corporation