Bayer Medical Care Recalls
9 recalls on record · September 12, 2018 to June 10, 2026 · Medical Devices
Recalls by year
18
19
20
21
23
26
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
June 10, 2026Medical DevicesFDA
Bayer Medical Care, Inc. recalls Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad¿ Avanta Injection System
Hazard: Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.
January 18, 2023Medical DevicesFDA
Bayer Medical Care, Inc. recalls Source Administration Set (SAS) used with the Medrad Intego PET Infusion System, Sterile Catalog number: INT CSS Affec
Hazard: Black specks may be present on the packaged needles
November 10, 2021Medical DevicesFDA
Bayer Medical Care, Inc. recalls The MEDRAD Twist & Go Disposable Syringe 150 mL with Quick Fill Tube is a syringe made from a clear polycarbonate materi
Hazard: Labeling error: MEDRAD Mark 7 Arterion Injection System Syringes (ART 700 SYR) labeled as Twist & Go Syringes (TAG 150 SYR)
December 2, 2020Medical DevicesFDA
Bayer Medical Care, Inc. recalls Brand Name: MEDRAD Stellant FLEX Syringe Kits Common Name: Angiographic Syringe Catalog/REF Number: FLEXD-150-SCS
Hazard: Specific lots of MEDRAD Stellant FLEX 150 mL syringe kits may have holes and/or cracks that have the potential to allow fluid to leak out of the syringe.
January 23, 2019Medical DevicesFDA
Bayer Medical Care, Inc. recalls MAVIG Portegra or Portegra2 Overhead Counterpoise System (OCS) used in conjunction with the Medrad Stellant CT Injection
Hazard: If the head mount knob is not properly tightened and the operator directly uses the injector head to move the unit, it may become disengaged, causing the injector head to become loose or fall.
October 17, 2018Medical DevicesFDA
Bayer Medical Care, Inc. recalls Codan Filter (Codan catalog number BC 693) used with Medrad Patient Administration Sets (PAS) and Intego PET Infusion Sy
Hazard: Bayer has received reports of loose particulate in the filter when the dust cap is removed. Bayer conducted further testing confirming the dust cap may generate particulate which may potentially enter the fluid path.
September 12, 2018Medical DevicesFDA
Bayer Medical Care, Inc. recalls Overhead Counterpoise Systems (OCS), OCS, 580mm CEILING MOUNT, PORTEGRA2, Catalog # OCS CEIL 58, Material Number 8443376
Hazard: The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.
September 12, 2018Medical DevicesFDA
Bayer Medical Care, Inc. recalls Overhead Counterpoise Systems (OCS) OCS CONF an accessory device for use with Bayer injector systems as a support for an
Hazard: The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.
September 12, 2018Medical DevicesFDA
Bayer Medical Care, Inc. recalls Overhead Counterpoise Systems (OCS), OCS CONF,STD CEIL MNT, STD HORZ ARM, Catalog # OCS 115A, Material Number 59880253
Hazard: The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.