RecallWatchUSA

Home / Recalls / Medical Devices

Bayer Medical Care, Inc. recalls Codan Filter (Codan catalog number BC 693) used with Medrad Patient Administration Sets (PAS) and Intego PET Infusion Sy

Recalled October 17, 2018 · FDA recall Z-0097-2019 · Bayer Medical Care

Hazard
Bayer has received reports of loose particulate in the filter when the dust cap is removed. Bayer conducted further testing confirming the dust cap may generate particulate which may potentially enter the fluid path. · FDA Class II
Units
1,142,836 (329,944 units in U.S.)
Sold at
Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, and WI. Foreign countries: UAE, Austria, Australia, Bosnia and Herzegovina, Belgium, Bahrain, Canada, China, Chile, Costa Rica, Czechia, Denmark, German, Spain, Estonia, Egypt, Finland, France, Great Britain, Croatia, Hungary, Israel, Italy, Jordan, Kuwait, Kazakhstan, Lebanon, Lithuania, Morocco, Malta, Mexico, Netherlands, Norway, Oman, Panama, Poland, Portugal, Romania, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, and Turkey

Description

Bayer has received reports of loose particulate in the filter when the dust cap is removed. Bayer conducted further testing confirming the dust cap may generate particulate which may potentially enter the fluid path.

Product
Codan Filter (Codan catalog number BC 693) used with Medrad Patient Administration Sets (PAS) and Intego PET Infusion Systems Product Usage: 1.2 micron filter for use with Intravenous Administration Sets
Manufacturer(s)
Bayer Medical Care, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Bayer Medical Care, Inc. recalls Codan Filter (Codan catalog number BC 693) used with Medrad Patient Administration Sets (PAS) and Intego PET Infusion Sy | RecallWatchUSA