Home / Brands / Beckman Coulter
Beckman Coulter Recalls
332 recalls on record · August 29, 2012 to August 12, 2026 · Medical Devices
Recalls by year
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
August 12, 2026Medical DevicesFDA
Beckman Coulter Inc. recalls REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) on the following instruments: - B1018-280 MicroScan
Hazard: Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
August 12, 2026Medical DevicesFDA
Beckman Coulter Inc. recalls REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018
Hazard: Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
April 29, 2026Medical DevicesFDA
Beckman Coulter, Inc. recalls Access Total T4 Calibrator, Catalog No. 33805
Hazard: Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.
April 15, 2026Medical DevicesFDA
Beckman Coulter, Inc. recalls MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use with MicroScan
Hazard: Due to manufacturing issue (incorrectly processing) with in vitro multidrug resistant test.
April 15, 2026Medical DevicesFDA
Beckman Coulter, Inc. recalls MicroScan Neg MIC 3J REF C54814
Hazard: Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics
April 1, 2026Medical DevicesFDA
Beckman Coulter, Inc. recalls MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining ant
Hazard: Due to the likely presence of contamination in well(s).
February 11, 2026Medical DevicesFDA
Beckman Coulter, Inc. recalls CellMek SPS Sample Preparation System, REF: C44603, with software
Hazard: Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.
February 11, 2026Medical DevicesFDA
Beckman Coulter, Inc. recalls COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjun
Hazard: Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.
January 14, 2026Medical DevicesFDA
Beckman Coulter, Inc. recalls Beckman Coulter UniCel Dxl 800, Part numbers: 973100, A71456, A71457, A25288; Immunoassay Method, Troponin Subunit
Hazard: Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs, the instrument will be unable to complete initialization and transition to Ready mode and the instrument will not process patient samples, QCs or calibrators. This causes delayed results.
January 14, 2026Medical DevicesFDA
Beckman Coulter, Inc. recalls Beckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin Subunit
Hazard: Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs, the instrument will be unable to complete initialization and transition to Ready mode and the instrument will not process patient samples, QCs or calibrators. This causes delayed results.
December 17, 2025Medical DevicesFDA
Beckman Coulter, Inc. recalls Access 2 Reaction Vessels, individual, disposable, plastic tubes used to process patient, quality control and calibratio
Hazard: It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could cause causing instrument errors (e.g., pipetting, rake, wash carousel errors) that could delay patient results.
November 26, 2025Medical DevicesFDA
Beckman Coulter Inc. recalls DxFLEX Flow Cytometer Catalog Number C78500 UDI-DI code: 15099590754013 The DxFLEX Flow Cytometer is intended for us
Hazard: Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.
November 12, 2025Medical DevicesFDA
Beckman Coulter Inc. recalls SYNCHRON Systems Direct LDL Cholesterol (LDLD) Reagent, REF: 969706
Hazard: Beckman Coulter has identified that there is an increased probability Synchron Systems Direct LDL Cholesterol (LDLD REF 969706) reagent lot M404166 may generate false high results. The issue was discovered internally and confirmed on August 19, 2025 by Beckman Coulter.
November 12, 2025Medical DevicesFDA
Beckman Coulter, Inc. recalls Beckman Coulter UniCel DxI 800, Part Numbers: 973100, A71456; Photometric Chemistry Analyzer
Hazard: An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
November 12, 2025Medical DevicesFDA
Beckman Coulter, Inc. recalls Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer
Hazard: An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
November 5, 2025Medical DevicesFDA
Beckman Coulter, Inc. recalls CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU
Hazard: Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.
October 29, 2025Medical DevicesFDA
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Part Number C11137
Hazard: Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.
October 29, 2025Medical DevicesFDA
Beckman Coulter, Inc. recalls DxC 500i Clinical Analyzer Modules, DxC 500 AU Module w/ISE, DxC 500i (Part Number C63522), DxC 500 AU Module, DxC 500i
Hazard: Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.
October 15, 2025Medical DevicesFDA
Beckman Coulter Inc. recalls SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only.
Hazard: Specific SYNCHRON Systems Modular Chemistry Phosphorus (PHOSm REF 467868) reagent lots may fail calibration before the lot expiration date which causes a delay in obtaining patient results.
October 8, 2025Medical DevicesFDA
Beckman Coulter Inc. recalls DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part
Hazard: On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.
September 24, 2025Medical DevicesFDA
Beckman Coulter Inc. recalls UniCel DxH 600 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF:B23858
Hazard: Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
September 24, 2025Medical DevicesFDA
Beckman Coulter Inc. recalls UniCel DxH 800 COULTER Cellular Analysis System, configured with the Led HGB Photometer REFs:B24802, 629029, B63322
Hazard: Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
September 24, 2025Medical DevicesFDA
Beckman Coulter Inc. recalls UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520
Hazard: Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
September 24, 2025Medical DevicesFDA
Beckman Coulter Inc. recalls UniCel DxH 900 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C11478
Hazard: Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
August 20, 2025Medical DevicesFDA
Beckman Coulter, Inc. recalls Access Thyroglobulin assay, a simultaneous one-step immunoenzymatic assay, Catalog Number 33860
Hazard: Access Thyroglobulin reagent lot 439163 may generate erroneously high patient results. Falsely increased Access Thyroglobulin results may lead a physician to pursue unnecessary diagnostic imaging studies and/or inappropriate therapy adjustments in patients being monitored for residual or recurrent thyroid cancer.
August 13, 2025Medical DevicesFDA
Beckman Coulter, Inc. recalls DxI 600 Access Immunoassay Analyzer W/Spot B, Part Number A71460
Hazard: Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
August 13, 2025Medical DevicesFDA
Beckman Coulter, Inc. recalls DxI 600 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71461
Hazard: Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
August 13, 2025Medical DevicesFDA
Beckman Coulter, Inc. recalls UniCel DxI 600 Access Immunoassay Analyzer, Part Number A30260
Hazard: Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
August 13, 2025Medical DevicesFDA
Beckman Coulter, Inc. recalls DxI 800 Access Immunoassay Analyzer W/Spot B, Part Number A71456
Hazard: Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
August 13, 2025Medical DevicesFDA
Beckman Coulter, Inc. recalls DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457
Hazard: Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.