RecallWatchUSA

Home / Recalls / Medical Devices

Beckman Coulter, Inc. recalls CellMek SPS Sample Preparation System, REF: C44603, with software

Recalled February 11, 2026 · FDA recall Z-1235-2026 · Beckman Coulter

Hazard
Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management. · FDA Class II
Units
50
Sold at
Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.

Description

Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.

Product
CellMek SPS Sample Preparation System, REF: C44603, with software
Manufacturer(s)
Beckman Coulter, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Beckman Coulter, Inc. recalls CellMek SPS Sample Preparation System, REF: C44603, with software | RecallWatchUSA