January 6, 2021Medical DevicesFDA
Becton Dickinson & Co. recalls BD Phoenix" PMIC-109: IVD is intended for the in vitro rapid identification (ID) of gram-positive bacteria from pure cu
Hazard: BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
January 6, 2021Medical DevicesFDA
Becton Dickinson & Co. recalls BD Phoenix PMIC/ID-109: IVD is intended for the in vitro rapid identification (ID) of gram-positive bacteria from pure
Hazard: BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
January 6, 2021Medical DevicesFDA
Becton Dickinson & Co. recalls BD Phoenix PMIC-108: IVD is intended for the in vitro rapid identification (ID) of gram-positive bacteria from pure cul
Hazard: BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
January 6, 2021Medical DevicesFDA
Becton Dickinson & Co. recalls BD Phoenix PMIC/ID-108: IVD is intended for the in vitro rapid identification (ID) of gram-positive bacteria from pure
Hazard: BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
January 6, 2021Medical DevicesFDA
Becton Dickinson & Co. recalls BD Phoenix PMIC-107: IVD is intended for the in vitro rapid identification (ID) of gram-positive bacteria from pure cul
Hazard: BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
January 6, 2021Medical DevicesFDA
Becton Dickinson & Co. recalls BD Phoenix PMIC/ID-107: IVD is intended for the in vitro rapid identification (ID) of gram-positive bacteria from pure
Hazard: BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
January 6, 2021Medical DevicesFDA
Becton Dickinson & Co. recalls BD Phoenix PMIC/ID-105: IVD is intended for the in vitro rapid identification (ID) of gram-positive bacteria from pure
Hazard: BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
December 23, 2020Medical DevicesFDA
Becton Dickinson & Company recalls BD PrecisionGlide Needle 30G x 1/2 (0.3 mm X 13mm) REF 305106 - Product Usage: used for general purpose injection and a
Hazard: BD has confirmed the presence of blood on the cannulas of the BD PrecisionGlide needles 30G X 1/2" Lot No. 9128897. The possibility of infection may be present if the blood is not noticed prior to use by the clinician.
December 9, 2020Medical DevicesFDA
Becton Dickinson & Company recalls BD Ultra-Fine Pen Needles 32G x 4mm (5/32' )-intended for parenteral administration of a drug from a cartridge contained
Hazard: Product shelf cartons incorrectly labeled as products intended for the Latin American market. Although the product description on the label 32G x 4mm pen needles is correct, information pertinent to users in the US regarding compatibility with specific pen needles is missing.
November 4, 2020Medical DevicesFDA
Becton Dickinson & Company recalls Bard 3F SL PowerMidline Full Kit CE-indicated for short term access to the peripheral venous system for selected intrave
Hazard: Kits do not contain the stylet/T-lock assembly indicated as a component on the label, may lead to a procedural delay and possibly discard of the kit.
November 4, 2020Medical DevicesFDA
Becton Dickinson & Company recalls Bard 3F SL PowerMidline Basic Kit CE-indicated for short term access to the peripheral venous system for selected intrav
Hazard: Kits do not contain the stylet/T-lock assembly indicated as a component on the label, may lead to a procedural delay and possibly discard of the kit.
October 28, 2020Medical DevicesFDA
Becton Dickinson & Company recalls BD Nexiva Closed IV Catheter System Dual Port (20 GA 1.75 IN) Catalog (Ref) No. 383538 BD Nexiva closed IV cathete
Hazard: Dull/blunt needles within the IV Catheter System may lead to missed insertion attempts and delay in diagnosis or treatment until a new needle is inserted
September 2, 2020Medical DevicesFDA
Becton Dickinson & Company recalls BD SafetyGlideTM 3ml 25G x 1 3mL SafetyGlide syringes, 50 units in pack
Hazard: Printed scale markings on the 3mL SafetyGlide syringes, with catalog # 305924, Lot# 8289599, are printed incorrectly. The scale is skewed to different degrees, resulting in missing and/or partially printed scale numbers and scale lines.
August 12, 2020Medical DevicesFDA
Becton Dickinson & Co. recalls BD MAX SARS-CoV-2 Reagents, Cat. No. 445003 (UDI 00382904450038) and 445003-01 (UDI (01)603829044500301) - Product Usage
Hazard: BD is experiencing elevated rates of potential false positive results from certain customers when using the BD SARS-CoV-2 Reagents for the BD MAX System. False positive test results could lead to a delay in diagnosis and potentially cause the patient to be exposed to COVID-19 if they are isolated with an infected individual.
July 29, 2020Medical DevicesFDA
Becton Dickinson & Company recalls SafeStep Huber Needle Set; Catalog Numbers LH-0029 (UDI 00801741066160), LH-0030 (UDI 00801741066184), LH-0031 (UDI 008
Hazard: Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
July 29, 2020Medical DevicesFDA
Becton Dickinson & Company recalls PowerLoc MAX Power Injectable Infusion Set; Catalog Numbers 0131910 (UDI 00801741047329), 0131915 (UDI 00801741047336),
Hazard: Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
July 29, 2020Medical DevicesFDA
Becton Dickinson & Company recalls SafeStep Huber Needle Set with Y Injection Site and Medegen Needleless Injection Cap; Catalog Numbers LH-0029YN (UDI 008
Hazard: Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
July 29, 2020Medical DevicesFDA
Becton Dickinson & Company recalls SafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access Kit; Catalog Numbers PA-0031G (UDI
Hazard: Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
July 29, 2020Medical DevicesFDA
Becton Dickinson & Company recalls SafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing; Catalog Number CPA00040 (UDI 00801741143892)
Hazard: Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
July 29, 2020Medical DevicesFDA
Becton Dickinson & Company recalls SafeStep Huber Needle Set Port Access Kit; Catalog Numbers PA-0029 (UDI 00801741085482), PA-0029YN (UDI 00801741085499),
Hazard: Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
July 29, 2020Medical DevicesFDA
Becton Dickinson & Company recalls SafeStep Huber Needle Set myPICK Port Access Kit; Catalog Numbers CPA00020A (UDI 00801741129827), CPA00042A (UDI 0080174
Hazard: Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
July 29, 2020Medical DevicesFDA
Becton Dickinson & Company recalls SafeStep Huber Needle Set AllPoints Port Access System; Catalog Numbers AN152075 (UDI 00801741119484), AN152075G (UDI 00
Hazard: Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
July 29, 2020Medical DevicesFDA
Becton Dickinson & Company recalls PowerLoc MAX Power Injectable Infusion Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access Kit; Catalog Numb
Hazard: Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
July 29, 2020Medical DevicesFDA
Becton Dickinson & Company recalls PowerLoc MAX Power Injectable Infusion Port Access Kit; Catalog Numbers 2131910 (UDI 00801741044151), 2131915 (UDI 00801
Hazard: Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
July 29, 2020Medical DevicesFDA
Becton Dickinson & Company recalls PowerLoc MAX Power-Injectable Infusion Set myPICK Port Access Kit; Catalog Numbers CPA00001 (UDI 00801741098345), CPA000
Hazard: Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
July 29, 2020Medical DevicesFDA
Becton Dickinson & Company recalls PowerLoc MAX Power Injectable Infusion Set AllPoints Port Access System; Catalog Numbers AN132075 (UDI 00801741118968),
Hazard: Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
May 27, 2020Medical DevicesFDA
Becton Dickinson & Company recalls BARD MYPICC KIT 5F Dual-Lumen, Catalog Number CK000435
Hazard: The kits contain surgical gowns which were subject to a recall by the supplier.
May 27, 2020Medical DevicesFDA
Becton Dickinson & Company recalls BARD MYPICC KIT 4F Dual-Lumen, Catalog Number CK000418B
Hazard: The kits contain surgical gowns which were subject to a recall by the supplier.
May 27, 2020Medical DevicesFDA
Becton Dickinson & Company recalls BARD MYPICC KIT 3F Single-Lumen, Catalog Number CK000417B
Hazard: The kits contain surgical gowns which were subject to a recall by the supplier.
May 27, 2020Medical DevicesFDA
Becton Dickinson & Company recalls BARD MYPICC KIT 5F Triple-Lumen, Catalog Number CK000278
Hazard: The kits contain surgical gowns which were subject to a recall by the supplier.