January 22, 2025Medical DevicesFDA
Boston Scientific Corporation recalls ALTRUA 2 DR Pacemaker, SL (Model Number S702) and EL (Model Number S722)
Hazard: A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
January 22, 2025Medical DevicesFDA
Boston Scientific Corporation recalls ESSENTIO DR Pacemaker, SL (Model Number L101, SL MRI (Model Number L111), EL (Model Number L121), and EL MRI (Model Numb
Hazard: A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
January 22, 2025Medical DevicesFDA
Boston Scientific Corporation recalls PROPONENT DR Pacemaker, SL (Model Number L201), (VDD) SL (Model Number L209), SL MRI (Model Number L211), EL (Model Numb
Hazard: A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
January 22, 2025Medical DevicesFDA
Boston Scientific Corporation recalls ACCOLADE DR Pacemaker, SL (Model Number L301), SL MRI (Model Number L311), EL (Model Number L321), and EL MRI (Model Num
Hazard: A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
December 4, 2024Medical DevicesFDA
Boston Scientific Corporation recalls AMS Artificial Urinary Sphincter Pressure Regulating Balloon 61-70 cm H2O
Hazard: Potential for mislabeled packaging in which a device labeled as a Pressure Regulating Balloon (UPN 72400024) which instead contained a Control Pump (UPN 72400098)
December 4, 2024Medical DevicesFDA
Boston Scientific Corporation recalls AMS Artificial Urinary Sphincter Control Pump
Hazard: Potential for mislabeled packaging in which a device labeled as a Pressure Regulating Balloon (UPN 72400024) which instead contained a Control Pump (UPN 72400098)
November 20, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific SMARTFREEZE CRYO CONSOLE: Console: Material Number M004CRBS4000; CONSOLE REFURB: Material Number
Hazard: Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
November 20, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific POLARx FIT BALLOON CATHETER LT US, Material Number M004CRBS2160
Hazard: Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
November 20, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific POLARx FIT BALLOON CATHETER LT OUS, Material Number M004CRBS2110
Hazard: Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
November 20, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific POLARx FIT BALLOON CATHETER ST US, Material Number M004CRBS2060
Hazard: Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
November 20, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific POLARx FIT BALLOON CATHETER ST OUS, Material Number M004CRBS2010
Hazard: Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
November 20, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific POLARx BALLOON CATHETER LT 28MM OUS, Material Number M004CRBS2100
Hazard: Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
November 20, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific POLARx BALLOON CATHETER ST 28MM OUS, Material Number M004CRBS2000.
Hazard: Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
October 30, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific myLUX Patient Application for an Apple Mobile device, Model Number 2939, Version 2.0.101, 2.0.110, ca
Hazard: Patients using the Model 2939 myLUX Patient Application on their Apple mobile device (e.g., iPhone) may not be transmitting data from their Insertable Cardiac Monitor to the LATITUDE Clarity website.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO MOB UPGRADE US; the system utilizes an int
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO MOB UPGRADE JAPAN; the system utilizes an
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO MOB UPGRADE EU ROW; the system utilizes an
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO INT UPGRADE US; the system utilizes an int
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO INT UPGRADE JAPAN; the system utilizes an
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO + MOB Z UPGRADE JP; the system utilizes an
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO + INT Z UPGRADE JP; the system utilizes an
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM US; the system utilizes an interactive dis
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM JAPAN; the system utilizes an interactive
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM EVAL US; the system utilizes an interactiv
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM EVAL EU + ROW; the system utilizes an inte
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM EVAL - ZERO COST US; the system utilizes a
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM EU + ROW; the system utilizes an interacti
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM - ZERO COST US; the system utilizes an int
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM - ZERO COST JAPAN; the system utilizes an
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
October 23, 2024Medical DevicesFDA
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM - ZERO COST EU + RO; the system utilizes a
Hazard: Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.