January 24, 2024Medical DevicesFDA
Cook Medical Llc recalls MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429
Hazard: The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.
January 20, 2021Medical DevicesFDA
Cook Medical Llc recalls Disposable Humidification Flask, Reference Part Number K-MINC-CTS-S, Order Number G32707
Hazard: A small percentage of the Disposable Humidification Flasks may have a manufacturing defect that can cause the silicone stoppers to dislodge or gas to leak from the flasks, impeding gas flow to the incubation chambers of the Mini Incubator.