RecallWatchUSA

Home / Recalls / Medical Devices

Cook Medical Llc recalls MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429

Recalled January 24, 2024 · FDA recall Z-0722-2024 · Cook Medical

Hazard
The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient. · FDA Class II
Units
14 US; 101 OUS
Sold at
Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM

Description

The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.

Product
MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429
Manufacturer(s)
Cook Medical Llc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Cook Medical Llc recalls MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429 | RecallWatchUSA