Home / Brands / Dako Denmark A/S
Dako Denmark A/S Recalls
3 recalls on record · December 26, 2012 to May 27, 2020 · Medical Devices
Recalls by year
12
16
20
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
May 27, 2020Medical DevicesFDA
Dako Denmark A/S recalls Dako Autostainer instruments: Autostainer Link 48, Model AS480; Autostainer Plus Link Instrument, Model AS100; Autosta
Hazard: The Autostainer instrument may not have a syringe tray installed which could result in a small volume of excess buffer being applied to several slides. In addition to the syringe tray, the updated Basic User Guide will reduce the potential risk of false negative results in the case of leakage of buffer and/or reagents that may impact staining.
August 10, 2016Medical DevicesFDA
Dako Denmark A/S recalls EnVision FLEX+, Mouse, High pH (Link), Product No. K8002, K8004. Intended for use in immunohistochemistry together wi
Hazard: The pH of Target Retrieval Solution in the kit measures 4-5 instead of the correct pH of 9.1.
December 26, 2012Medical DevicesFDA
Dako Denmark A/S recalls CoverStainer Slide Rack The CoverStainer Slide Rack is an accessory to the CoverStainer instrument. The purpose is to
Hazard: The expected life time of the CoverStainer slide racks (CS 10330) does not meet established performance requirements, originating from variations in the manufacturing process.