RecallWatchUSA

Home / Recalls / Medical Devices

Dako Denmark A/S recalls Dako Autostainer instruments: Autostainer Link 48, Model AS480; Autostainer Plus Link Instrument, Model AS100; Autosta

Recalled May 27, 2020 · FDA recall Z-2074-2020 · Dako Denmark A/S

Hazard
The Autostainer instrument may not have a syringe tray installed which could result in a small volume of excess buffer being applied to several slides. In addition to the syringe tray, the updated Basic User Guide will reduce the potential risk of false negative results in the case of leakage of buffer and/or reagents that may impact staining. · FDA Class II
Units
796 devices
Sold at
Distribution was nationwide. There was government/military distribution. Foreign distribution was made to Canada, Australia, Bangaladesh, Belarus, Brazil, Bulgaria, Chile, China, Colombia, Croatia, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Macedonia, Morocco, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Poland, Qatar, Romania, Russia, Saudi Arabia, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, Uruguay, and Vietnam.

Description

The Autostainer instrument may not have a syringe tray installed which could result in a small volume of excess buffer being applied to several slides. In addition to the syringe tray, the updated Basic User Guide will reduce the potential risk of false negative results in the case of leakage of buffer and/or reagents that may impact staining.

Product
Dako Autostainer instruments: Autostainer Link 48, Model AS480; Autostainer Plus Link Instrument, Model AS100; Autostainer, Model S3400; and Autostainer Plus, Model S3800.
Manufacturer(s)
Dako Denmark A/S
Made in
Denmark

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Dako Denmark A/S recalls Dako Autostainer instruments: Autostainer Link 48, Model AS480; Autostainer Plus Link Instrument, Model AS100; Autosta | RecallWatchUSA