Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
January 22, 2020Medical DevicesFDA
Degania Silicone, Ltd. recalls Medline Thermistor Foley Catheter 14 FR 5-10 ml, non sterile Catalog Number: 102201101463MD {Medline code 55346) Mon
Hazard: Temperature deviation did not meet release specifications
August 21, 2019Medical DevicesFDA
Degania Silicone, Ltd. recalls Smith & Nephew Negative Pressure Wound Therapy 15FR Channel Wound Drain PN: 02872 An introduction/drainage catheter
Hazard: 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
August 21, 2019Medical DevicesFDA
Degania Silicone, Ltd. recalls Medline Silicone Drain 15Fr 3/16" Full Fluted PN: DYNJWE2188HB An introduction/drainage catheter is a device that
Hazard: 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
August 21, 2019Medical DevicesFDA
Degania Silicone, Ltd. recalls Cardinal Health Jackson-Pratt Channel Drain 19FR, 3/4 flutes, with trocar Cardinal Health: PN JP-2225 An introducti
Hazard: 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
August 21, 2019Medical DevicesFDA
Degania Silicone, Ltd. recalls Cardinal Health Jackson-Pratt Channel Drain 19FR, Full flutes, with trocar Cardinal Health: PN: JP-2191 An introduct
Hazard: 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
August 21, 2019Medical DevicesFDA
Degania Silicone, Ltd. recalls Cardinal Health Jackson-Pratt Channel Drain 19FR, Full flutes Cardinal Health: PN JP-2190 An introduction/drainage c
Hazard: 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
August 21, 2019Medical DevicesFDA
Degania Silicone, Ltd. recalls Cardinal Health Jackson-Pratt Channel Drain 15FR, Full flutes, with trocar Product Number: JP-2189 An introduct
Hazard: 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
August 21, 2019Medical DevicesFDA
Degania Silicone, Ltd. recalls Cardinal Health Jackson-Pratt Channel Drain 15FR, Full flutes Cardinal Health PN: JP-2188 An introduction/drainage
Hazard: 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
August 21, 2019Medical DevicesFDA
Degania Silicone, Ltd. recalls Cardinal Health Jackson-Pratt Channel Drain 15FR Cardinal PN:JP-2223 An introduction/drainage catheter is a devi
Hazard: 15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Temperature sensing catheter 16FR, Catalogue Number 102201101663NU Product Usage: This is sterile, single use, two-w
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Temperature sensing catheter 14FR, Catalogue Number 102201101463NU Product Usage: This is sterile, single use, two-w
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Temperature sensing catheter 18FR, Catalogue Number 102203101863DO Product Usage: This is sterile, single use, two-w
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Temperature sensing catheter 16FR, Catalogue Number 102203101663DO Product Usage: This is sterile, single use, two-w
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Temperature sensing catheter 14FR, Catalogue Number 102203101463DO Product Usage: This is sterile, single use, two-w
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Thermistor Foley catheter, nonsterile, Catalogue Number 102201101463MD Product Usage: This is sterile, single use, t
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Thermistor Foley catheter, nonsterile, Catalogue Number 102201100863MD Product Usage: This is sterile, single use, two
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Thermistor Foley catheter, nonsterile, Catalogue Number 102201101663MD Product Usage: This is sterile, single use, t
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Thermistor Foley catheter, nonsterile, Catalogue Number 102201101863MD Product Usage: This is sterile, single use, t
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101080MD Product Usage: This is st
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101680MD Product Usage: This is st
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101880MD Product Usage: This is st
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201101480MD Product Usage: This is st
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Temperature Sensing 100% Silicone Foley Catheter, sterile, Catalogue Number 102201100880MD Product Usage: This is st
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Foley Temperature Sensor Catheter 14FR, Catalogue Number 102201101480JP Product Usage: This is sterile, single use,
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Foley Temperature Sensor Catheter 12FR, Catalogue Number 102201101280JP Product Usage: This is sterile, single use,
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Foley Temperature Sensor Catheter 10FR, Catalogue Number 102201101080JP Product Usage: This is sterile, single use,
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Foley Temperature Sensor Catheter 8FR, Catalogue Number 102201100880JP Product Usage: This is sterile, single use, t
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Foley Catheter with temperature sensor 400TM 18FR, Catalogue Number 102201101880TY Product Usage: This is sterile, s
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Foley Catheter with temperature sensor 400TM 16FR, Catalogue Number 102201101680TY Product Usage: This is sterile, s
Hazard: The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
September 26, 2018Medical DevicesFDA
Degania Silicone, Ltd. recalls Foley Catheter with temperature sensor 400TM 14FR, Catalogue Number 102201101480TY Product Usage: This is sterile, s