Home / Recalls / Medical Devices
Degania Silicone, Ltd. recalls Temperature sensing catheter 16FR, Catalogue Number 102201101663NU Product Usage: This is sterile, single use, two-w
Recalled September 26, 2018 · FDA recall Z-2785-2018 · Degania Silicone
Hazard
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient. · FDA Class II
Units
1000
Sold at
Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
Description
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
Product
Temperature sensing catheter 16FR, Catalogue Number 102201101663NU Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
Manufacturer(s)
Degania Silicone, Ltd.
Made in
Israel
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →