DePuy Orthopaedics Recalls

155 recalls on record · July 11, 2012 to April 10, 2024 · Medical Devices

Recalls by year

12
13
14
15
16
17
18
19
20
21
22
23
24

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

July 29, 2015Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls ATTUNE RP CR ARTICULATION SURFACE SIZE 3-8 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the
Hazard: Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.
July 29, 2015Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls ATTUNE CONV RP CR ARTICULATION SURFACE SZ1, 2, 9, and 10 INTENDED USE: The Tibial Articulation Surface Trial snaps
Hazard: Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.
July 29, 2015Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls ATTUNE RP PS ARTICULATION SURFACE SIZE 3-8 INTENDED USE: The Tibial Articulation Surface Trial snaps together with t
Hazard: Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.
July 29, 2015Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls ATTUNE FB CR ARTICULATION SURFACE SIZE 3-8 INTENDED USE: The Tibial Articulation Surface Trial snaps together with t
Hazard: Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.
July 29, 2015Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls ATTUNE FB PS ARTICULATION SURFACE SIZES 3-8 INTENDED USE: The Tibial Articulation Surface Trial snaps together with
Hazard: Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.
June 3, 2015Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls LCS COMPLETE RPS inserts, various sizes. Knee prosthesis component for orthopedic surgery.
Hazard: The LCS COMPLETE RPS Knee System has higher rates of revision in Australia when the native patella is not resurfaced when compared to other unresurfaced PS implants. If the LCS COMPLETE RPS Knee System is implanted and the native patella is not resurfaced, the patient may experience patello-femoral pain, potentially requiring a secondary procedure.
June 3, 2015Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls LCS COMPLETE RPS Femoral Implants, various sizes and orientation. Knee prosthesis component for orthopedic surgery.
Hazard: The LCS COMPLETE RPS Knee System has higher rates of revision in Australia when the native patella is not resurfaced when compared to other unresurfaced PS implants. If the LCS COMPLETE RPS Knee System is implanted and the native patella is not resurfaced, the patient may experience patello-femoral pain, potentially requiring a secondary procedure.
April 8, 2015Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls RECLAIM Assembled Implant Inserter Adaptor orthopedic manual surgical instrument for hip implantation.
Hazard: This lot of the RECLAIM Assembled Implant Inserter Adaptor Instrument may fracture during surgery and there is the potential for plastic pieces to be left in the patient.
March 11, 2015Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls SMARTSET GHV Gentamicin Bone Cement. A self-curing, radiopaque, polymethyl methacrylate based cement containing antibio
Hazard: One lot of SMARTSET GHV Gentamicin Bone Cement is partially agglomerated . This may result in surgical delays while a replacement package of cement is retrieved and mixed.
January 21, 2015Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls The ATTUNE INTUITION Impaction Handle is a re-useable instrument utilized in knee replacement procedures. The instrument
Hazard: Complaints have indicated that the ATTUNE INTUITION Impaction Handle lever has fractured during impaction. If the lever fractures, there is the potential for fractured pieces of the device, or the inner spring, if released, to be left in the patient if not observed during surgery.
January 14, 2015Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls I.M. Hole Locator Instrument. Orthopedic manual surgical instrument. The IM Hole locator guides the IM drill dur
Hazard: Use of excessive force when impacting may lead to intra-operative femoral fracture.
January 14, 2015Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls Femoral Impactor ATTUNE INTUITION Impactor. The ATTUNE INTUITION Impactors are re-useable instruments utilized in knee
Hazard: ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient.
January 14, 2015Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls Rotating Tibial Platform ATTUNE INTUITION Impactor. The ATTUNE INTUITION Impactors are re-useable instruments utilized
Hazard: ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient.
January 14, 2015Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls Fixed Tibial bearing ATTUNE INTUITION Impactor The ATTUNE INTUITION Impactors are re-useable instruments utilized in kn
Hazard: ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient.
December 24, 2014Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls Ceramic Femoral Head. Intended for use in total hip arthroplasty applications to replace the articular surface of t
Hazard: Two complaints reported that the BIOLOX delta TS Ceramic Femoral Articul/eze12/14 36mm +12 Head (product code 1365-36-740) did not fit like the trial; the definite hip construct felt loose. The loosening refers to tissue tension and not the taper connection between the femoral head and stem. The investigations found that the small size sleeve (+1.5 femoral head offset) was incorrectly used in the
November 19, 2014Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls ATTUNE Intuition Distal Femoral Jig; sold non-sterile; product is wrapped in bubble wrap and then sealed inside a polyet
Hazard: Issuing a device correction because if the pin bushing is over loaded or inverted, there is the potential for partial or full displacement and fracture of the plastic around the pin bushing.
October 15, 2014Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls Charnley PN Retract & Handle, REF 962004000, Non-Sterile-DuPuy International Ltd. The Charnley Pin Retractor and Hand
Hazard: The small extraction peg of the Charnley Pin Retractor and Handle set is breaking from the pin.
October 15, 2014Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls RECLAIM Distal Reamer Extension Non Sterile REF 2975-00-500 Intended as an attachment device to the distal reamers
Hazard: Specific lots are being recalled due to the potential for the tabs to break, leaving fragments in the patient.
July 30, 2014Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls RECLAIM DISTAL TAPERED Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manua
Hazard: The product can be difficult to remove from the Distal Stem both out of the package and after proximal reaming.
July 30, 2014Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls RECLAIM DISTAL TAPERED, Angled Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthoped
Hazard: The product can be difficult to remove from the Distal Stem both out of the package and after proximal reaming.
July 30, 2014Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls RECLAIM Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the l
Hazard: The product can be difficult to remove from the Distal Stem both out of the package and after proximal reaming.
April 16, 2014Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls GLOBAL¿ Anchor Peg Glenoid Plus (APG+) 2.5MM Breakaway Guide Pin Part # 2230-00-019, Non-Sterile, DePuy, Johnson & Joh
Hazard: Urgent Device Correction notifications are being sent to USA distributors with follow-up to surgeons concerning updated surgical techniques stressing the possibility of the shoulder pin guide breaking and potentially being left in the patient.
March 19, 2014Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls REF 623421 RS-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Right , X-Small, Sterile R Qty 1
Hazard: DePuy Orthopaedics has identified the potential for holes to develop in the inner and outer flexible pouches that form the sterile barrier for S-ROM Noiles Rotating Hinge Femur with Pin devices. The outer carton and shrink wrap are intact. A package redesign is underway to resolve this issue. S-ROM Noiles Rotating Hinge Femur with Pin devices will remain in US Distributors inventory during the pa
March 19, 2014Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls REF 623421 L S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Left , X-Small, Sterile R Qty 1
Hazard: DePuy Orthopaedics has identified the potential for holes to develop in the inner and outer flexible pouches that form the sterile barrier for S-ROM Noiles Rotating Hinge Femur with Pin devices. The outer carton and shrink wrap are intact. A package redesign is underway to resolve this issue. S-ROM Noiles Rotating Hinge Femur with Pin devices will remain in US Distributors inventory during the pa
March 19, 2014Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls REF 623411 R S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Right, Small, Sterile R Qty 1
Hazard: DePuy Orthopaedics has identified the potential for holes to develop in the inner and outer flexible pouches that form the sterile barrier for S-ROM Noiles Rotating Hinge Femur with Pin devices. The outer carton and shrink wrap are intact. A package redesign is underway to resolve this issue. S-ROM Noiles Rotating Hinge Femur with Pin devices will remain in US Distributors inventory during the pa
March 19, 2014Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls REF 623411 L, S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Left, Small, Sterile R Qty 1
Hazard: DePuy Orthopaedics has identified the potential for holes to develop in the inner and outer flexible pouches that form the sterile barrier for S-ROM¿ Noiles Rotating Hinge Femur with Pin devices. The outer carton and shrink wrap are intact. A package redesign is underway to resolve this issue. S-ROM Noiles Rotating Hinge Femur with Pin devices will remain in US Distributors inventory during the p
March 19, 2014Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls REF 623401 R, S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Right, Medium, Sterile R Qty 1
Hazard: DePuy Orthopaedics has identified the potential for holes to develop in the inner and outer flexible pouches that form the sterile barrier for S-ROM¿ Noiles Rotating Hinge Femur with Pin devices. The outer carton and shrink wrap are intact. A package redesign is underway to resolve this issue. S-ROM Noiles Rotating Hinge Femur with Pin devices will remain in US Distributors inventory during the p
March 19, 2014Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls REF 623401 L S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Left, medium, Sterile R Qty 1
Hazard: DePuy Orthopaedics has identified the potential for holes to develop in the inner and outer flexible pouches that form the sterile barrier for S-ROM¿ Noiles Rotating Hinge Femur with Pin devices. The outer carton and shrink wrap are intact. A package redesign is underway to resolve this issue. S-ROM Noiles Rotating Hinge Femur with Pin devices will remain in US Distributors inventory during the
March 5, 2014Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT) Used during primary total knee arthroplasty
Hazard: Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This decrease in strength could cause implants to fracture under what would normally be acceptable loads.
December 4, 2013Medical DevicesFDA
DePuy Orthopaedics, Inc. recalls SIGMA¿ LCS¿ HIGH PERFORMANCE" MBT PUNCH NONCEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare
Hazard: DePuy Orthopaedics, Inc. is issuing a voluntary device correction of the HP M.B.T. Keel Punch Knee Instrument because of the potential for tabs on the connection end to fracture and be left in the patient.