Home / Brands / Fort Defiance Industries
Fort Defiance Industries Recalls
2 recalls on record · July 18, 2018 to October 27, 2021 · Medical Devices
Recalls by year
18
21
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
October 27, 2021Medical DevicesFDA
Fort Defiance Industries, LLC recalls P2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641
Hazard: The rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.
July 18, 2018Medical DevicesFDA
Fort Defiance Industries, LLC recalls FDI FORT DEFIANCE INDUSTRIES P2131 AUTOMATED FIELD STEAM STERILIZER, NSN: 6530-01-641-4641 Product Usage: The Fort D
Hazard: During long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system that can potentially interfere with the rear sight gauge and / or the water liquid level switch.