RecallWatchUSA

Home / Recalls / Medical Devices

Fort Defiance Industries, LLC recalls FDI FORT DEFIANCE INDUSTRIES P2131 AUTOMATED FIELD STEAM STERILIZER, NSN: 6530-01-641-4641 Product Usage: The Fort D

Recalled July 18, 2018 · FDA recall Z-2411-2018 · Fort Defiance Industries

Hazard
During long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system that can potentially interfere with the rear sight gauge and / or the water liquid level switch. · FDA Class II
Units
199 units
Sold at
US Nationwide Distribution in the states of PA. VA. TX

Description

During long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system that can potentially interfere with the rear sight gauge and / or the water liquid level switch.

Product
FDI FORT DEFIANCE INDUSTRIES P2131 AUTOMATED FIELD STEAM STERILIZER, NSN: 6530-01-641-4641 Product Usage: The Fort Defiance Industries Automated Field Steam Sterilizer Model P2131 is designed for sterilization of porous and non-porous, heat- and moisture-stable materials (e.g., surgical instruments and textiles) used in healthcare facilities. The P2131 sterilizer is a transportable device designed for use in a variety of austere environments.
Manufacturer(s)
Fort Defiance Industries, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Fort Defiance Industries, LLC recalls FDI FORT DEFIANCE INDUSTRIES P2131 AUTOMATED FIELD STEAM STERILIZER, NSN: 6530-01-641-4641 Product Usage: The Fort D | RecallWatchUSA