GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING Recalls

41 recalls on record · February 15, 2023 to January 28, 2026 · Medical Devices

Recalls by year

23
24
25
26

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

January 28, 2026Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls GE Healthcare Omni Legend, sold under the following names: OMNI Legend 16, Omni Legend 21, OMNI Legend 32, Omni Legen
Hazard: There is a potential intermittent issue on certain Omni Legend systems that can result in a streaking artifact in the PET clinical scan images. This streaking artifact is most easily identified in transaxial slices on the acquisition console (both corrected and non-corrected for attenuation).
January 21, 2026Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls MAXXUS, System, Tomography, Computed, Emission
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
January 21, 2026Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls SMV BodyTrack, System, Tomography, Computed, Emission
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
January 21, 2026Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls SMV DSI, System, Tomography, Computed, Emission
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
January 21, 2026Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls SMV DSX New Line, System, Tomography, Computed, Emission
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
October 29, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Starcam/Camstar; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
October 29, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Sopha Medical Vision (SMV) DSXi/DST-XLi,/ DST-XL/ DSTi; dual-head nuclear medicine gamma camera designed specifically fo
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
October 29, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Optima NX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
October 29, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Millenium MPS; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
October 29, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Millenium MPR; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
October 29, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Millenium Myosight; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
October 29, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Millenium MG/MC; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
October 29, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Maxicam; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
October 29, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Elscint Helix; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
October 29, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Elscint Varicam; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
October 29, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Elscint SPX6; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
October 29, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Elscint SPX4; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
October 29, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Elscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
October 29, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Elscint Cardial; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Hazard: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
July 16, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System
Hazard: Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
July 16, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye 4 Nuclear Medicine system
Hazard: Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
January 22, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls GE Healthcare Varicam Millennium VG Discovery VH, Model/Catalog Numbers: 1) ASM000085; 2) ASM001092; 3) SYS000001
Hazard: GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
January 22, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls GE Healthcare Infinia II Hawkeye 4, Model/Catalog Numbers: 1) H2799PP; 2) H2799PR; 3) H2799YC; 4) H2799YD; 5)
Hazard: GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
January 22, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls GE Healthcare Infinia I Hawkeye 1, Model/Catalog Numbers: 1) H3000WN; 2) H3000YM; 3) H3000YS; System, Tomography,
Hazard: GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
January 22, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls GE Healthcare Infinia II, Model/Catalog Numbers: 1) H2799ZP; 2) H3000WT; System, Tomography, Computed, Emission
Hazard: GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
January 22, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls GE Healthcare Optima NM/CT 640, Model/Catalog Numbers: 1) H3100JZ; System, Tomography, Computed, Emission
Hazard: GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
January 22, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission
Hazard: GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
January 22, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls GE Healthcare Discovery NM/CT 670 ES, Model/Catalog Numbers: 1) H3100XF ; System, Tomography, Computed, Emission
Hazard: GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
January 22, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls GE Healthcare Discovery NM/CT 670 PRO, Model/Catalog Numbers: 1) 5376204-70-54-2; 2) H2401AK; 3) H2402AC; 4) H3
Hazard: GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
January 22, 2025Medical DevicesFDA
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls GE Healthcare Discovery NM/CT 670 DR, Model/Catalog Numbers: 1) 5376204-70-56-2; 2) H2401AW; 3) H3100AN; 4) H31
Hazard: GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.