GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System

Recalled July 16, 2025 · FDA recall Z-2098-2025 · GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING

Hazard
Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury. · FDA Class II
Units
21
Sold at
Worldwide - US Nationwide distribution.

Description

Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.

Product
Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System
Manufacturer(s)
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Made in
Israel