Handicare AB Recalls
3 recalls on record · October 18, 2017 to October 21, 2020 · Medical Devices
Recalls by year
17
20
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
October 21, 2020Medical DevicesFDA
Handicare AB recalls SwiftHook sold as the following: a. Individually as SwiftHook b. Included with Vega505EE mobile lifts Product Usage:
Hazard: Carry bar may detach during use of mobile lifts and ceiling lifts equipped with the Swifthook component.
May 20, 2020Medical DevicesFDA
Handicare AB recalls EVA Floor Lifts, Model Numbers: 400, 450, 600
Hazard: After feedback from customers and internal suspicion, Handicare has become aware that abnormal use of Eva-lifts could lead to a deterioation in the joint of the liftarm. This could lead to injury if the joint breaks. All lifts must therefore be checked and reported back to the manuafacturer.
October 18, 2017Medical DevicesFDA
Handicare AB recalls Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009
Hazard: A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.