RecallWatchUSA

Home / Recalls / Medical Devices

Handicare AB recalls Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009

Recalled October 18, 2017 · FDA recall Z-0014-2018 · Handicare AB

Hazard
A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye. · FDA Class II
Units
9 units
Sold at
OH, PA, MD

Description

A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.

Product
Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009
Manufacturer(s)
Handicare AB
Made in
Sweden

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov