March 9, 2022Medical DevicesFDA
Lacrimedics Inc recalls Collagen Plugs, CP3, 0.3mm, CP4, 0.4mm, CP5, 0.5 mm, Sterile, EO, Rx Only, CE Dry eye treatment
Hazard: The sterile pouch seal may contain channels that could affect the seal integrity and potentially compromise sterility.
March 9, 2022Medical DevicesFDA
Lacrimedics Inc recalls OPAQUE Herrick Lacrimal Plugs, HP3, 0.3 mm, HP5, 0.5mm, HP7, 0.7.mm, STERILE, EO, Rx only, CE Dry eye treatment
Hazard: The sterile pouch seal may contain channels that could affect the seal integrity and potentially compromise sterility.
February 18, 2015Medical DevicesFDA
Lacrimedics Inc recalls VisiPlugST for the Lacrimal Efficiency Test. It is also sold in the EU as CollaSyn Plugs. Model number 1813, 1814, 181
Hazard: Moisture level may cause the VisiPlugST (or CollaSyn Plugs) to break apart, or crumble upon contact. The firm's data also suggests that a 4 year shelf life may not be sustained.