Home / Recalls / Medical Devices
Lacrimedics Inc recalls VisiPlugST for the Lacrimal Efficiency Test. It is also sold in the EU as CollaSyn Plugs. Model number 1813, 1814, 181
Recalled February 18, 2015 · FDA recall Z-1081-2015 · Lacrimedics
Hazard
Moisture level may cause the VisiPlugST (or CollaSyn Plugs) to break apart, or crumble upon contact. The firm's data also suggests that a 4 year shelf life may not be sustained. · FDA Class III
Units
781 units
Sold at
Worldwide Distribution-US (nationwide) including the states of AR, CA, FL, ID, IL, KS, MA, MI, NC, NJ, NY, OR, PA, and TX and to the countries of Australia, Brazil, Czech Republic, France, Ireland, Mexico, Peru, Switzerland, and United Kingdom.
Description
Moisture level may cause the VisiPlugST (or CollaSyn Plugs) to break apart, or crumble upon contact. The firm's data also suggests that a 4 year shelf life may not be sustained.
Product
VisiPlugST for the Lacrimal Efficiency Test. It is also sold in the EU as CollaSyn Plugs. Model number 1813, 1814, 1815, 1903, 1904, and 1905. This device is cylindrical in shape, approximately 1.75-2.00mm in length, and available in three sizes: 0.3mm, 0.4mm and 0.5mm. Plugs are place in the upper and/or lower horizontal canaliculus. Two plugs are placed into a foam carrier and nitrogen sealed inside a foil pouch. The foil pouch is sent to sterilizer for gamma sterilization. Two sterile pouches are packaged in a box with an IFU.
Manufacturer(s)
Lacrimedics Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →