RecallWatchUSA

Home / Brands / LEICA BIOSYSTEMS NUSSLOCH

LEICA BIOSYSTEMS NUSSLOCH Recalls

10 recalls on record · May 1, 2024 to March 5, 2025 · Medical Devices

Recalls by year

24
25

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

March 5, 2025Medical DevicesFDA
LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Brand Name: Leica Biosystems Cryostats Product Name: CM1950 Model/Catalog Number: 1491950C4US Software Version: firmw
Hazard: Firm became aware of software issues related to most recent firmware update which could result in loss of patient tissue.
October 23, 2024Medical DevicesFDA
LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica Biosystems Cryostat Model Number CM3050 S, Material Numbers: a) 14047033518, b) 14903050S01, c) 14903050S02, d) 1
Hazard: As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
October 23, 2024Medical DevicesFDA
LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica Biosystems Cryostat Model Number CM1950, Material Numbers: a) 14047742467, b) 1491950C1US, c) 1491950C2US, d) 149
Hazard: As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
October 23, 2024Medical DevicesFDA
LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c) 14901900700, d) 149
Hazard: As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
October 23, 2024Medical DevicesFDA
LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica Biosystems Cryostat Model Number CM1850, Material Numbers: a) 14047131148, b) 14901850001, c) 14901850001R0501,
Hazard: As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
October 23, 2024Medical DevicesFDA
LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica Biosystems Cryostat Model Number CM1510 S, Material Numbers: a) 491510C002, b) 1491510C003; for freezing and sect
Hazard: As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
October 23, 2024Medical DevicesFDA
LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue samples
Hazard: As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
August 14, 2024Medical DevicesFDA
LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica HistoCore Arcadia H, REF: 14039354100 and 14039357258, Paraffin Embedding Station.
Hazard: An issue with safe usage of the device was identified whereby toxic smoke and internal fire developed resulting from a problem associated with the sealing of the paraffin tank.
June 12, 2024Medical DevicesFDA
LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica HistoCore PEGASUS Tissue Processor, REF 14048858005.
Hazard: Poorly processed and/or damaged biopsy tissue specimens on the device resulting from reagent levels exceeding the maximum fill level marks on reagent bottles or in the paraffin tanks.
May 1, 2024Medical DevicesFDA
LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica HistoCore PEGASUS (PROCESSOR, TISSUE, AUTOMATED)
Hazard: There is an issue regarding poorly processed and/or damaged biopsy tissue specimens on the HistoCore PEGASUS / HistoCore PEGASUS Plus resulting from an incorrect assignment of the carryover value set for created or edited protocols.