RecallWatchUSA

Home / Recalls / Medical Devices

LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c) 14901900700, d) 149

Recalled October 23, 2024 · FDA recall Z-0088-2025 · LEICA BIOSYSTEMS NUSSLOCH

Hazard
As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019. · FDA Class II
Units
25 units
Sold at
US Nationwide distribution.

Description

As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.

Product
Leica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c) 14901900700, d) 1491900UVU2; for freezing and sectioning tissue samples
Manufacturer(s)
LEICA BIOSYSTEMS NUSSLOCH GMBH
Made in
Germany

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c) 14901900700, d) 149 | RecallWatchUSA