Linvatec Corp. Recalls

21 recalls on record · August 15, 2012 to July 17, 2013 · Medical Devices

Recalls by year

12
13

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

July 17, 2013Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls ***REF C7122***QTY 1***APEX Arthroscopy Tubing Set. Intended for use for joint irrigation and dissention during arthr
Hazard: Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and C7122 APEX Arthroscopy Tubing Set, Two Connection, contained a scratch/grooved area on the inside diameter of the pressure fitting that may cause the APEX Irrigation (pump) Console to sense a pressure drop and increase fluid flow to the surgical site.
July 17, 2013Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls ***REF C7120***APEX Arthroscopy Tubing Set. Intended for use for joint irrigation and dissention during arthroscopic
Hazard: Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and C7122 APEX Arthroscopy Tubing Set, Two Connection, contained a scratch/grooved area on the inside diameter of the pressure fitting that may cause the APEX Irrigation (pump) Console to sense a pressure drop and increase fluid flow to the surgical site.
July 10, 2013Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls GENESYS Cross FT Suture Anchor with Two #2 (5 metric) Hi-FI Registered Trade Mark Sutures Rx Only STERILE EO CONMED LINV
Hazard: Incorrect blue/white suture was used.
May 29, 2013Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls Ref: VP1600 HD Still Capture System ConMed Linvatec . Captures, stores and transfers images generated by a medical
Hazard: The VP1600 HD Still Capture Image System is recalled due to an anomaly between the software code in the VP1600 console and the communications interface for various keyboard key strokes (Caps lock and Num Lock) which may lock up and requires the user to reboot to continue.
March 13, 2013Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls D3000I Advantage Drive System, 230v, surgical instrument motor. Product Usage: The Advantage Drive Systems function
Hazard: The Membrane Switch Panel (referred to as "front panel" of the component part number A34-003-207, used on both the D3000 and D3000I Advantage Drive System, contained the incorrect symbology per third edition IEC60606-01 Medical Electrical Equipment: General requirements for basic safety and essential performance.
March 13, 2013Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls D3000 Advantage Drive System, 115v Product Usage: The Advantage Drive Systems function as a powered instrument syst
Hazard: The Membrane Switch Panel (referred to as "front panel" of the component part number A34-003-207, used on both the D3000 and D3000I Advantage Drive System, contained the incorrect symbology per third edition IEC60606-01 Medical Electrical Equipment: General requirements for basic safety and essential performance.
February 13, 2013Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls ConMed Linvatec Biomaterials, PopLok bone Punch PKL-35M, 3.5mm Orthopedic manual surgical instrument - Bone Punch
Hazard: ConMed Linvatech recalled one lot of their PopLok bone Punch 3.5mm because the metal shaft diameter is 4.5mm when it should be 3.5mm.
November 28, 2012Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls "***SENTINEL***Cannulated Drill Bit***S8585***8.5 X 229mm (9")***CONMED LINVATEC 11311 CONCEPT BLVD LARGO, FL 33773-4908
Hazard: Linvatec Corporation dba ConMed Linvatec in Largo, FL is recalling the S8580, 8.0MM SENTINEL Cannulated Drill Bit, lot numbers 322225 and 322226; and the S8585, 8.5MM SENTINEL Cannulated Drill Bit, lot number 322287. Recall was initiated due to the incorrect labeling on package.
November 28, 2012Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls "***SENTINEL***Cannulated Drill Bit***S8580***8 X 229mm (9")***CONMED LINVATEC 11311 CONCEPT BLVD LARGO, FL 33773-4908 U
Hazard: Linvatec Corporation dba ConMed Linvatec in Largo, FL is recalling the S8580, 8.0MM SENTINEL Cannulated Drill Bit, lot numbers 322225 and 322226; and the S8585, 8.5MM SENTINEL Cannulated Drill Bit, lot number 322287. Recall was initiated due to the incorrect labeling on package.
November 7, 2012Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls "***All in one CrossFT***Punch***PFT-00M***4.5mm to 6.5mm***CONMED LINVATEC 11311 CONCEPT BLVD LARGO, FL 33773-4908 USA*
Hazard: ConMed Linvatec is recalling the product due to the punch being mis-assembled with the incorrect handled. The correct handle is a clear anodized silver colored handle.
October 24, 2012Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls NP211H***2009-02***CONMED LINVATEC 11311 CONCEPT BLVD LARGO, FL 33773-4908 USA*** To reattach soft tissue to bone in
Hazard: Linvatec Corporation dba ConMed Linvatec in Largo, FL is recalling PressFT HI-FI Sutures (items NP211, NP212, NP261, NP262, NP211H, NP261H-all lot numbers) due to the potential for the distal tip of the driver to become separated from the driver shaft after successful insertion of the implant.
October 24, 2012Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls "***PressFT***2.1 w/ One #2 (5 metric) Hi-Fi***Suture***NP211***2009-02***CONMED LINVATEC 11311 CONCEPT BLVD LARGO, FL 3
Hazard: Linvatec Corporation dba ConMed Linvatec in Largo, FL is recalling PressFT HI-FI Sutures (items NP211, NP212, NP261, NP262, NP211H, NP261H-all lot numbers) due to the potential for the distal tip of the driver to become separated from the driver shaft after successful insertion of the implant.
October 24, 2012Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls "***PressFT***2.1 w/ Two #0 (3.5 metric) Hi-Fi***Sutures***NP212***2009-02***CONMED LINVATEC 11311 CONCEPT BLVD LARGO, F
Hazard: Linvatec Corporation dba ConMed Linvatec in Largo, FL is recalling PressFT HI-FI Sutures (items NP211, NP212, NP261, NP262, NP211H, NP261H-all lot numbers) due to the potential for the distal tip of the driver to become separated from the driver shaft after successful insertion of the implant.
October 24, 2012Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls "***Press FT***2.6 w/ One #2 (5 metric) Hi-Fi***Suture, HIP***N2***2009-02***CONMED LINVATEC 11311 CONCEPT BLVD LARGO, F
Hazard: Linvatec Corporation DBA ConMed Linvatec in Largo, FL is recalling Press FT HI-FI Sutures (items NP 211, NP 212, NP 261, NP 262, NP211H, NP261H-all lot numbers) due to the potential for the distal tip of the driver to become separated from the driver shaft after successful insertion of the implant.
October 24, 2012Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls "***PressFT***2.6 w/ One #2 (5 metric) Hi-Fi***Suture***NP261***2009-02***CONMED LINVATEC 11311 CONCEPT BLVD LARGO, FL 3
Hazard: Linvatec Corporation dba ConMed Linvatec in Largo, FL is recalling PressFT HI-FI Sutures (items NP211, NP212, NP261, NP262, NP211H, NP261H-all lot numbers) due to the potential for the distal tip of the driver to become separated from the driver shaft after successful insertion of the implant.
October 24, 2012Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls "***PressFT***2.6 w/ Two #1 (4 metric) Hi-Fi***Sutures***NP262***2009-02***CONMED LINVATEC 11311 CONCEPT BLVD LARGO, FL
Hazard: Linvatec Corporation dba ConMed Linvatec in Largo, FL is recalling PressFT HI-FI Sutures (items NP211, NP212, NP261, NP262, NP211H, NP261H-all lot numbers) due to the potential for the distal tip of the driver to become separated from the driver shaft after successful insertion of the implant.
October 3, 2012Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls Catheter Irrigation Set labeled in part: ***24K Arthroscopy Outflow/Suction Tubing Set*** This Arthroscopy Tubing Set
Hazard: ConMed Linvatec recalled their 24k Arthroscopy Outflow/Suction Tubing Set (Product Number 24k100)because there is a possibility that the product may have a breach in the Tyvek seal that could potentially compromise the sterility of the contents.
October 3, 2012Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls ***REF LS7700***Qty 1*** EC REP*** ConMed Linvatec Xenon Universal Light Source
Hazard: Linvatec Corporation dba ConMed Linvatec, Largo, FL initiated a recall of their products LS7700 Xenon Universal Light Source and LS7600 Xenon Light Source are intended to be used with an endoscope to provide illumination during endoscopic procedures.
October 3, 2012Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls ***REF LS7600***Qty 1*** EC REP*** ConMed Linvatec Xenon Light Source
Hazard: Linvatec Corporation dba ConMed Linvatec, Largo, FL initiated a recall of their products LS7700 Xenon Universal Light Source and LS7600 Xenon Light Source are intended to be used with an endoscope to provide illumination during endoscopic procedures.
August 15, 2012Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls PopLok Punch labeled in part: "PKL-35M***3.5mm***QTY 1***PopLok Punch***ConMed Linvatec Biomaterials, Ltd. Hermiankatu
Hazard: ConMed Linvatec recalled one lot of their PopLok bone Punch 3.5MM because the metal shaft diameter is 4.5mm when it should have been 3.5mm.
August 15, 2012Medical DevicesFDA
Linvatec Corp. dba ConMed Linvatec recalls CrossFT Punch, Product Number PFT-45M. Manufacturer: CONMED LINVATEC. The intended use is to create a hole to facilita
Hazard: ConMed Linvatec recalled lot number #156920, PFT-45M, CrossFT Punch because there is a possibility that the device was etched incorrectly listing the item number as PFT-00M and includes incorrect depth lines for insertion.