Linvatec Corp. dba ConMed Linvatec recalls Ref: VP1600 HD Still Capture System ConMed Linvatec . Captures, stores and transfers images generated by a medical

Recalled May 29, 2013 · FDA recall Z-1340-2013 · Linvatec Corp.

Hazard
The VP1600 HD Still Capture Image System is recalled due to an anomaly between the software code in the VP1600 console and the communications interface for various keyboard key strokes (Caps lock and Num Lock) which may lock up and requires the user to reboot to continue. · FDA Class II
Units
267 units
Sold at
Worldwide Distribution-USA (nationwide) including the Virgin Islands and the states of CA, CT, FL, GA, IN, KY, KS, LA, MD, MI, MN, MO, MS, MT, NE, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, and WI, and the countries of Austria, Belgium, Canada, Denmark, Hong Kong and Zambia.

Description

The VP1600 HD Still Capture Image System is recalled due to an anomaly between the software code in the VP1600 console and the communications interface for various keyboard key strokes (Caps lock and Num Lock) which may lock up and requires the user to reboot to continue.

Product
Ref: VP1600 HD Still Capture System ConMed Linvatec . Captures, stores and transfers images generated by a medical endoscopic camera system during surgical procedures.
Manufacturer(s)
Linvatec Corp. dba ConMed Linvatec