February 12, 2025Medical DevicesFDA
LivaNova USA, Inc. recalls SenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for Vagus Nerve Stimul
Hazard: Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
February 12, 2025Medical DevicesFDA
LivaNova USA, Inc. recalls SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimula
Hazard: Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
February 5, 2025Medical DevicesFDA
LivaNova USA, Inc. recalls Brand Name: SenTiva DUO" Product Name: VNS Therapy¿ SenTiva DUO", Model 1000-D Model/Catalog Number: 1000-D Software
Hazard: Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.
February 5, 2025Medical DevicesFDA
LivaNova USA, Inc. recalls Brand Name: SenTiva" Product Name: VNS Therapy" SenTiva", Model 1000 Model/Catalog Number: 1000 Software Version: NA
Hazard: Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.
February 9, 2022Medical DevicesFDA
LivaNova USA, Inc. recalls Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mm electrode
Hazard: A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience discomfort or pain due to delivery of stimulation to an unintended location. If the inner lead conductor is exposed to body fluids the lead may break causing loss of therapy.
December 16, 2020Medical DevicesFDA
LivaNova USA Inc recalls VNS Therapy SENTIVA DUO # 1000-D - Product Usage: indicated for the treatment of chronic or recurrent depression in pati
Hazard: During internal testing, it was found that upon a device reset, the Generator exhibits the incorrect Model Number when interrogated.
November 25, 2020Medical DevicesFDA
LivaNova USA Inc recalls VNS Therapy¿ AspireSR¿ Generator
Hazard: Due to an extra digit being inadvertently added to the serial number of a subset of implantable pulse generators, they become not compatible with therapy programmers with software version 11.0.4.
July 29, 2020Medical DevicesFDA
LivaNova USA Inc. recalls LivaNova USA Sterile Disposable Connector,1/2x1/2, 20 units per case
Hazard: Some packages of sterile disposable connectors contained a 3/8x1/4 reducer instead of the specified 1/2x1/2 connector. No significant risk to health is expected as the issue is easily detectable and the different shape would prevent use of the connector.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Vein Irrigation Cannula with blunt tip. ASY VIC, STERILE 25/CS REF/ GTIN for Insert Label: VIC/ 00803622133107
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Vein Irrigation Cannula with blunt tip. ASY BIC, STERILE 25/CS REF/ GTIN for Insert Label:BIC/ 00803622133091
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls 1/2 - 3/8 REDUC CONN, STRL REF/ GTIN for Insert Label:EC2145S/ 00803622129322
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls 1/4X3/8 REDUC CONN, STRL REF/ GTIN for Insert Label:EC2140S/ 00803622129315
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls 3/8 STR CONN, STRL REF/ GTIN for Insert Label:EC2130S/ 00803622129292
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls 3/8X3/8X3/8 Y CONN, STRL REF/ GTIN for Insert Label:EC2105S/ 00803622129285
Hazard: XXX
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls 1/4X3/16X3/16YCONN,STRL REF/ GTIN for Insert Label:EC2102S/ 00803622129278
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Sterile Disposable Connectors ASY 1/2 CONN W/LL STRLE 24 REF/ GTIN for Insert Label:050608000/ 00803622123382
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Sterile Disposable Connectors ASY RDC Y 1/2X3/8X3/8 STR 24 REF/ GTIN for Insert Label:050529000/ 00803622123351
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Sterile Disposable Connectors ASY Y CONN 3/8 X 3/8 X 3/8 24 REF/ GTIN for Insert Label:050526000/ 00803622123320
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Sterile Disposable Connectors ASY Y CONN 1/4 X 1/4 X 1/4 24 REF/ GTIN for Insert Label:050524000/ 00803622123306
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Sterile Disposable Connectors ASY CONN EQUAL 3/16 WYE 24 REF/ GTIN for Insert Label:050523000/ 00803622123290
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Sterile Disposable Connectors ASY CONN REDUCER 1/2 X 3/8 24 REF/ GTIN for Insert label:050518000/ 00803622123283
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Sterile Disposable Connectors ASY CONN REDUCER 3/8 X 1/4 24 REF/ GTIN for Insert Label:050516000/ 00803622123276
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Sterile Disposable Connectors ASY S/1/4 TO 3/16 RED 24 REF/ GTIN for Insert Label:050514000/00803622123269
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Sterile Disposable Connectors ASY ASY STR CONN 3/8 STRLE 24 REF/ GTIN for Insert Label:050506000/ 00803622123238
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Sterile Disposable Connectors ASY S CONN 3/16X1/8 RDR ANLD24 REF/ GTIN for Insert Label:050513000/ 00803622123252
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Sterile Disposable Connectors ASY STR CONN 1/4 STRLE 24. REF/ GTIN for Insert Label:050504000/ 00803622123221
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Sterile Disposable Connectors ASY STR CONN 3/16 STRLE 24. REF/ GTIN for Insert Label:050503000/ 00803622123214
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Sterile Disposable Connectors ASY 1/4 X 3/8 REDUCER W/LL 24. REF/ GTIN for Insert Label:050434000/ 00803622123207
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Sterile Disposable Connectors ASY 3/16 X 3/16 STR CONNW/LL24. REF/ GTIN for Insert Label:050432000 / 00803622123184
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
July 8, 2020Medical DevicesFDA
LivaNova USA Inc. recalls Sterile Disposable Connectors ASY CONN Y 1/2X1/2X3/8 STR 24. REF/ GTIN for Insert Label:029165000/ 00803622123153
Hazard: The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.