Home / Recalls / Medical Devices
LivaNova USA, Inc. recalls SenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for Vagus Nerve Stimul
Recalled February 12, 2025 · FDA recall Z-1073-2025 · LivaNova USA
Hazard
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device. · FDA Class II
Units
48
Sold at
US: AK, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PN, SC, SD, TN, TX, VA, VT, WA, WI, WV, UT. OUS: AE, AR, AT, BE, BG, BR, CA, CL, CO, CR, CZ, DE, DK, GR, HR, IL, IN, IT, JO, KW, LB, MK, NL, OM, PA, PE, PL, QA, RO, RS, SA, SE, SG, SI, TR, TT, UK, ZA.
Description
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
Product
SenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for Vagus Nerve Stimulation (VNS).
Manufacturer(s)
LivaNova USA, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →