4 recalls on record · August 20, 2014 to August 20, 2014 · Medical Devices
August 20, 2014Medical DevicesFDA
Medical Positioning Inc recalls Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1219. Product Usage: Use
Hazard: Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.
August 20, 2014Medical DevicesFDA
Medical Positioning Inc recalls Medical Positioning Rapid Response HUT table with Dual 2 Way Drop Section, Powered Medical Examination table, Model # 12
Hazard: Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.
August 20, 2014Medical DevicesFDA
Medical Positioning Inc recalls Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1019. Product Usage: Use
Hazard: Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.
August 20, 2014Medical DevicesFDA
Medical Positioning Inc recalls Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1017. Product Usage: Used
Hazard: Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.