Medical Positioning Inc recalls Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1019. Product Usage: Use

Recalled August 20, 2014 · FDA recall Z-2209-2014 · Medical Positioning

Hazard
Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction. · FDA Class II
Units
2 beds
Sold at
Worldwide Distribution - US Nationwide in the states of CA, GA, FL, IL, IN, KS, KY, MD, MN, MO, MS, NC, NY, OH, TN, TX, VA and WI.

Description

Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.

Product
Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1019. Product Usage: Used to conduct syncope examinations.
Manufacturer(s)
Medical Positioning Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov