Normand-Info S.A.S.U Recalls
3 recalls on record · March 11, 2020 to June 16, 2021 · Medical Devices
Recalls by year
20
21
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
June 16, 2021Medical DevicesFDA
Normand-Info S.A.S.U. recalls Normand Remisol Advance Data Manager is a software device that interfaces between laboratory information systems (LIS) a
Hazard: Instructions For Use for software versions 1.5 to 1.9 did not provide sufficient instructions regarding the proper programming and testing of rules. Therefore, rules written by users that were not validated or incorrectly validated may lead to erroneous results.
August 12, 2020Medical DevicesFDA
Normand-Info S.A.S.U. recalls Remisol Advance running Software Version 1.12 The Normand Info Remisol Advance Data Manager collects and manages data
Hazard: Results from repeated run for WBC (White Bloodcell Count), UWBC (Urine Whit Bloodcell Count) and PLT (Platelets) are deleted in RADV. Use may lead to erroneous results.
March 11, 2020Medical DevicesFDA
Normand-Info S.A.S.U. recalls Remisol Advance running on hardware with Windows XP, Windows 7, Windows Server 2003 and Windows Server 2008 operating sy
Hazard: A patch was released for a critical remote code execution vulnerability in remote desktop services. This vulnerability can be exploited remotely without authentication on certain operating systems. An attacker who successfully exploited this vulnerability could execute arbitrary code on the target system; then install programs; view, change, or delete data; or create new accounts with full user rights. If the vulnerability could be exploited locally, it would lock down the computer.