Normand-Info S.A.S.U. recalls Normand Remisol Advance Data Manager is a software device that interfaces between laboratory information systems (LIS) a
Recalled June 16, 2021 · FDA recall Z-1831-2021 · Normand-Info S.A.S.U
Hazard
Instructions For Use for software versions 1.5 to 1.9 did not provide sufficient instructions regarding the proper programming and testing of rules. Therefore, rules written by users that were not validated or incorrectly validated may lead to erroneous results. · FDA Class II
Units
8,402 units with the affected software
Sold at
Worldwide distribution - US Nationwide distribution in the states of Alabama, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Indiana, Iowa, Kansas, Louisiana, Maine, Maryland, Massachusetts, Michigan, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, Pennsylvania, Rhode Island, South Carolina, Texas, West Virginia and Wyoming. The countries of Australia, Austria, Belgium, Brunei, Canada, China, Colombia, Congo, Czech Republic, Denmark, Ethiopia, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Lithuania, Malaysia, Mayotte, Morocco, Netherlands, Netherlands Antillean, New Zealand, Norway, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, UAE, UK, and Zimbabwe.
Description
Instructions For Use for software versions 1.5 to 1.9 did not provide sufficient instructions regarding the proper programming and testing of rules. Therefore, rules written by users that were not validated or incorrectly validated may lead to erroneous results.
Product
Normand Remisol Advance Data Manager is a software device that interfaces between laboratory information systems (LIS) and laboratory instruments.
Manufacturer(s)
Normand-Info S.A.S.U.
Made in
France