PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Recalls

209 recalls on record · February 23, 2022 to August 19, 2026 · Medical Devices

Recalls by year

22
23
24
25
26

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

June 5, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura Xper FD10 system with an ADN7NT patient table. Model Number: 722010, Software Version Number 8.1.100.
Hazard: When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
June 5, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012, Software Version Number 8.1.100.
Hazard: When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
February 28, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surg
Hazard: Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
February 28, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 5 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surg
Hazard: Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
February 28, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 3 with a Certeray generator -To perform image guidance in diagnostic, interventional, and minimally invasive sur
Hazard: Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
February 28, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the DIMMs (Dual In-line Memory Modules) used
Hazard: Dual In-Line Memory Modules (DIMMs) may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
February 28, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected
Hazard: Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
February 28, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of affected
Hazard: Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
February 28, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected
Hazard: Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
February 28, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of affected
Hazard: Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
February 7, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)
Hazard: Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.
February 7, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)
Hazard: Loss of connectivity between the FlexArm and the Table due to a software issue.
February 7, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion system, Interventional fluoroscopic x-ray system. System Product Name - Model Numbers Azurion 3M12 -7
Hazard: System may exhibit a loss of imaging functionality and data due to an issue which can cause the system to continuously restart. If the issue occurs during a procedure, there may be a delay and/or abortion of the procedure.
December 27, 2023Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion systems with monoplane fixed ceiling mounts: 722078 Azurion 7 M12 722079 Azurion 7 M20 722223 Azurion
Hazard: Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose
December 27, 2023Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura systems with monoplane fixed ceiling mounts: 722001 Allura Xper FD10C 722003 Allura Xper FD10 722006
Hazard: Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose
September 20, 2023Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura Xper, Allura Centron, Allura CV, Azurion, UNIQ Clarity, and MultiDiagnost-Eleva systems
Hazard: A damaged foot switch may cause instances where no or only intermittent x-ray radiation initiation is possible.
September 13, 2023Medical DevicesFDA
Philips Medical Systems Nederland B.V. recalls MultiDiagnost-Eleva
Hazard: A foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation
September 13, 2023Medical DevicesFDA
Philips Medical Systems Nederland B.V. recalls Azurion
Hazard: A foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation
September 13, 2023Medical DevicesFDA
Philips Medical Systems Nederland B.V. recalls Allura Xper series
Hazard: A foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation
September 6, 2023Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Wireless Foot Switch distributed with Philips Allura Xper and Azurion Interventional Fluoroscopic X-Ray Systems
Hazard: Loss of availability of the wireless foot switch during procedures.
August 9, 2023Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 M20. X-ray Image guided Therapy System.
Hazard: The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.
August 9, 2023Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10 and FD20. X-ray Imaging System.
Hazard: The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.
June 7, 2023Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/Model Number: (1) A
Hazard: Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure
April 26, 2023Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls 2D Perfusion, release R1.0.x, R1.1.x, R1.2, and R1.2.1
Hazard: There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.
April 26, 2023Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls SmartPerfusion, release R1.0, R1.1, R1.1.1, R1.1.5, R1.1.5.1 and R1.1.6
Hazard: There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.
May 11, 2022Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion systems with software release R1.x
Hazard: In the Azurion system, the user can add a new study to a patient by selecting the option Add Study . The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is initiated by pressing Start Procedure , the Patient Type changes inadvertently to a Patient Type different than the one selected as shown in the Table below. Patient type is one of the factors involved in the dose control process. The incorrect patient type changes the technique factors to be used by the system without notification to the user.
February 23, 2022Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Veradius Unity, Model #718132
Hazard: Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.
February 23, 2022Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Zenition 70, Model #718133
Hazard: Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.
February 23, 2022Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Zenition 50, Model #718096
Hazard: Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.