Home / Brands / Philips North America
Philips North America Recalls
530 recalls on record · April 5, 2017 to July 1, 2026 · Medical Devices
Recalls by year
17
18
19
20
21
22
23
24
25
26
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
October 20, 2021Medical DevicesFDA
Philips North America LLC recalls Philips, HeartStart FR2 + Defibrillator, Model # M3860A and M3861A
Hazard: Customers were not notified of previous recalls associated with various defibrillator models.
September 22, 2021Medical DevicesFDA
Philips North America Llc recalls The Azurion series (within the limits of the used Operation Room table) are intended for use to perform: "Image guidanc
Hazard: Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.
September 22, 2021Medical DevicesFDA
Philips North America Llc recalls The Allura Xper series is intended for use on human patients to perform: "Vascular, cardiovascular and neurovascular im
Hazard: Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.
September 22, 2021Medical DevicesFDA
Philips North America Llc recalls The Allura Xper series is intended for use on human patients to perform: "Vascular, cardiovascular and neurovascular ima
Hazard: Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.
September 15, 2021Medical DevicesFDA
Philips North America LLC recalls Philips, HEARTSTART, INFANT/CHILD - SMART PADS CARTRIDGE, REF: M5072A, (0-8 YEARS)< 55 lbs / 25 kg, CE 0123, NON-STERILE
Hazard: Infant/Child defibrillator pads contain incorrect labeling. This could lead to a delay in therapy, or delivery of lower energy than indicated for an adult patient
September 15, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia Ambition X- Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system c
Hazard: When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
September 15, 2021Medical DevicesFDA
Philips North America Llc recalls SmartPath to dStream for XR and 3.0T-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic d
Hazard: When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
September 15, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia 3.0T CX-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can
Hazard: When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
September 15, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia Elition X-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system c
Hazard: When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
September 15, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia Elition S-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system c
Hazard: When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
September 15, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia 3.0T-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can pr
Hazard: When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
September 15, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia Ambition S-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system
Hazard: When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
September 15, 2021Medical DevicesFDA
Philips North America Llc recalls SmartPath to dStream for 1.5T-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device.
Hazard: When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
September 15, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia 1.5T CX-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can
Hazard: When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
September 15, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia 1.5T S-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can
Hazard: When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
September 15, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia 1.5T Evolution-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The sys
Hazard: When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
September 15, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia 1.5T- a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can p
Hazard: When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
September 1, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia Elition S-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system ca
Hazard: Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury
September 1, 2021Medical DevicesFDA
Philips North America Llc recalls Achieva 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can pro
Hazard: Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury
September 1, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia CX 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can
Hazard: Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury
September 1, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia Elition X-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system ca
Hazard: Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury
September 1, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can pro
Hazard: Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury
September 1, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia 1.5T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can pro
Hazard: Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury
September 1, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia 1.5T Evoluton- Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The syste
Hazard: Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury
August 25, 2021Medical DevicesFDA
Philips North America Llc recalls 728333 Spectral CT 7500 -Computed Tomography X-ray system
Hazard: When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.
August 25, 2021Medical DevicesFDA
Philips North America Llc recalls 728332 IQon Spectral CT-Computed Tomography X-ray system
Hazard: When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.
August 25, 2021Medical DevicesFDA
Philips North America Llc recalls 728327 Ingenuity CT Upgrades-Computed Tomography X-ray system
Hazard: When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.
August 25, 2021Medical DevicesFDA
Philips North America Llc recalls Model 728321/728323 CT5000 Ingenuity Plus/Pro/Premium-Computed Tomography X-ray system
Hazard: When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.
August 25, 2021Medical DevicesFDA
Philips North America Llc recalls 728306 Brilliance iCT-Computed Tomography X-ray systems 728305 Brilliance iCT Upgrades
Hazard: When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.
July 7, 2021Medical DevicesFDA
Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728306 - Brilliance iCT 728311 - Brillia
Hazard: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.