March 21, 2018Medical DevicesFDA
Randox Laboratories recalls Immunoassay Premium Controls (3x4x5mL and 12x5mL packages)
Hazard: The ACTH in Immunoassay Premium and Premium Plus quality control does not meet the quoted reconstituted stability claim in the product IFU.
March 21, 2018Medical DevicesFDA
Randox Laboratories, Limited recalls Liquid Cardiac Control CQ5053
Hazard: Randox has confirmed that Troponin T and Troponin I in Liquid Cardiac quality control CQ5053 lot 4069CK does not meet the shelf-life claim in the product labelling. As the control may now fall outside the provided control ranges the product is to be recalled from the field.
January 3, 2018Medical DevicesFDA
Randox Laboratories Ltd. recalls Rx series HbA1c REF/Catalog Number HA3830, GTIN 05055273203561, IVD; -- RANDOX LABORATORIES LTD. , This product is suit
Hazard: Randox confirmed low calibration absorbance with HbA1c, Catalog Number HA3830, Batch 414553. The reagent may not continue to meet its specific performance characteristics throughout the shelf life of the product.
June 7, 2017Medical DevicesFDA
Randox Laboratories recalls Liquid Urine Control Level 3 For Professional Use in the quality control monitoring of In-Vitro Diagnostic Urinary Cl
Hazard: According to Randox Laboratories an investigation indicates a labelling error for 2 of the test methods listed for quantification of Total Protein (urine). The values for the mean of all instruments are incorrect by a factor of 10. Instrument specific values are correct. This does not affect the performance or stability claims of the product. As control results will be outside of the quoted range this may result in a delay in reporting the sample test results. However as the correct instrument specific means are also included in the value sheet this delay is unlikely. Customers have been referred to their Medical Director for further advice
December 21, 2016Medical DevicesFDA
Randox Laboratories recalls The Randox Liquid Cardiac Controls Level 1, Level 2, and Level 3 are liquid controls containing BNP, CK MB Mass, Digoxin
Hazard: An internal real-time stability monitoring program revealed decreased recovery of myoglobin (-20% bias) for Randox Immunoturbidimetric and Siemens Centaur Classic methods and CKMB (-21% bias) for Roche e411, BioMerieux Vidas, and Siemens Centaur Classic methods.
December 14, 2016Medical DevicesFDA
Randox Laboratories, Limited recalls Magnesium on RX Imola analyser IVD
Hazard: According to the firm, Carry over was observed when the amylase or pancreatic amylase test is run directly before or after the Magnesium assay. An update is required to the RX Imola analyser running order and an acid wash recommended when setting up the magnesium assay. A correction was made to the IFU for the Magnesium assay and all RX Imola customers in the USA were contacted with the updated instruction.
October 26, 2016Medical DevicesFDA
Randox Laboratories Ltd. recalls Randox Lipase (Colorimetric) Product Usage: A lipase test system is a device intended to measure the activity of the
Hazard: According to the firm carry over was observed when the Lipase test is run directly before or after the LDL and Triglycerides assays. Customers are instructed to amend the running order for all instruments using these reagents to ensure Lipase is not place directly before or after LDL or Triglycerides. A correction was made to the IFU for the Lipase assay and all customers who purchased the Lipase kit in the USA were contacted with the updated instruction.
October 12, 2016Medical DevicesFDA
Randox Laboratories Ltd. recalls IgE Calibrator Series
Hazard: According to the firm, the IgE Calibrator Series IE2492 (lot 387-392IE) value assignment has now been aligned to the 3rd International Standard for human serum IgE 11/234. Turbidimetric (Non-IFCC Cal.) Quality Control Targets and Ranges now apply. The calibrator value sheet is to be updated with the new values.