RecallWatchUSA

Home / Recalls / Medical Devices

Randox Laboratories Ltd. recalls Randox Lipase (Colorimetric) Product Usage: A lipase test system is a device intended to measure the activity of the

Recalled October 26, 2016 · FDA recall Z-0123-2017 · Randox Laboratories

Hazard
According to the firm carry over was observed when the Lipase test is run directly before or after the LDL and Triglycerides assays. Customers are instructed to amend the running order for all instruments using these reagents to ensure Lipase is not place directly before or after LDL or Triglycerides. A correction was made to the IFU for the Lipase assay and all customers who purchased the Lipase kit in the USA were contacted with the updated instruction. · FDA Class II
Units
18
Sold at
US Nationwide Distribution

Description

According to the firm carry over was observed when the Lipase test is run directly before or after the LDL and Triglycerides assays. Customers are instructed to amend the running order for all instruments using these reagents to ensure Lipase is not place directly before or after LDL or Triglycerides. A correction was made to the IFU for the Lipase assay and all customers who purchased the Lipase kit in the USA were contacted with the updated instruction.

Product
Randox Lipase (Colorimetric) Product Usage: A lipase test system is a device intended to measure the activity of the enzyme lipase in serum and plasma. Lipase measurements are used in the diagnosis and treatment of diseases of the pancreas such as acute pancreatitis and obstruction of the pancreatic duct.
Manufacturer(s)
Randox Laboratories Ltd.
Made in
Ireland

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Randox Laboratories Ltd. recalls Randox Lipase (Colorimetric) Product Usage: A lipase test system is a device intended to measure the activity of the | RecallWatchUSA