RecallWatchUSA

Home / Brands / Randox Laboratories

Randox Laboratories Recalls

98 recalls on record · October 12, 2016 to January 29, 2025 · Medical Devices

Recalls by year

16
17
18
19
20
21
22
23
24
25

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

January 29, 2025Medical DevicesFDA
Randox Laboratories, Limited recalls Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers. Catalo
Hazard: Lipoprotein (a) on Roche Cobas c501 with nmol/l claims does not have a 510(k) or PMA.
January 29, 2025Medical DevicesFDA
Randox Laboratories, Limited recalls Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers. Catal
Hazard: Lipoprotein (a) on Roche Cobas c501 with nmol/l claims does not have a 510(k) or PMA.
July 31, 2024Medical DevicesFDA
Randox Laboratories Ltd. recalls RX Series CYSC- Cystatin C Reagent -Intended for in vitro diagnostic (IVD) use as Randox Cystatin C assay on automated a
Hazard: Cystatin C Reagent marketed without a 510 (k)
May 15, 2024Medical DevicesFDA
Randox Laboratories Ltd. recalls Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug molecule and
Hazard: Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This will affect the testing and may delay in reporting results.
May 15, 2024Medical DevicesFDA
Randox Laboratories Ltd. recalls Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 c
Hazard: Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods.
April 24, 2024Medical DevicesFDA
Randox Laboratories Ltd. recalls Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426
Hazard: Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other method may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.
April 24, 2024Medical DevicesFDA
Randox Laboratories Ltd. recalls Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567
Hazard: Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.
April 3, 2024Medical DevicesFDA
Randox Laboratories Ltd. recalls Cystatin C Calibrator - In vitro diagnostic (IVD) use in the calibration of the Randox Cystatin C assay on automated ana
Hazard: Cystatin C Calibrator Series CVS 2699, is running with a negative bias compared to other methods
March 27, 2024Medical DevicesFDA
Randox Laboratories Ltd. recalls Liquid Protein Calibrators. C3 and Haptoglobin liquid protein calibrators used with Randox IgA, IgG, and IgM assays that
Hazard: The firm has realigned C3 and Haptoglobin in Liquid Protein Calibrators, IT2691, to reference material ERM-DA470k/IFCC.
February 14, 2024Medical DevicesFDA
Randox Laboratories Ltd. recalls RX Series Copper (Cu) Assay Ref. Number CU2340
Hazard: Randox Laboratories have released an update to the Carryover Avoidance Technical Bulletin to detail that Copper, should not be run directly after Total Protein, on the RX series instruments testing order. If copper assay is run after total protein it could lead to falsely elevated copper results.
January 24, 2024Medical DevicesFDA
Randox Laboratories Ltd. recalls Calibration Serum Level 3 - IVD For use as a Calibrator in clinical chemistry assays Catalog Number: CAL2351
Hazard: Target for Alkaline Phosphatase (ALP) for the AMP optimized to IFCC 37¿C method has been mis-assigned for the RX Series instruments by approximately 10% on Quality Control and patient samples
January 24, 2024Medical DevicesFDA
Randox Laboratories Ltd. recalls Calibration Serum Level 3 CAL2351
Hazard: CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.
January 24, 2024Medical DevicesFDA
Randox Laboratories Ltd. recalls Calibration Serum Level 3. CAL 3, REF CAL2351. Human Assayed Multi-Sera Level 2 (HN1530) Human Assayed Multi-Sera Level
Hazard: There have been transcription errors on the Instructions For Use (IFU) of the Calibration Serum Level 3 (CAL2351 lot 1260UE, 1262UE and 1315UE) for target values of Bilirubin and Triglycerides, on the IFU of Human Assayed Multi-Sera Level 2 (HN1530) for TBIC taget value and on IFU of Human Assayed Multi-Sera Level 3 (HE 1532) for Lipase target value.
January 10, 2024Medical DevicesFDA
Randox Laboratories Ltd. recalls SEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use with in vitro assays for determination of IgM antibodies to C
Hazard: A decrease in the reactivity of HSV Type 1/2 lgM in the Serology ToRCH lgM Positive Control when tested on the DiaSorin Liaison XL.
December 13, 2023Medical DevicesFDA
Randox Laboratories Ltd. recalls RANDOX Total Bilirubin.
Hazard: Randox has had reports of elevated patient results using Plasma (lithium heparin) samples with Total Bilirubin reagent (TBil) Vanadate Oxidation Method.
November 8, 2023Medical DevicesFDA
Randox Laboratories Ltd. recalls RX Monaco-IVD Clinical Chemistry Analyser Catalog Number: RX5000
Hazard: RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
November 8, 2023Medical DevicesFDA
Randox Laboratories Ltd. recalls RX Modena-IVD Clinical Chemistry Analyser Catalog Number: RX9000
Hazard: RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
November 8, 2023Medical DevicesFDA
Randox Laboratories Ltd. recalls RX Imola-IVD Clinical Chemistry Analyzer Catalog Number: RX4900. RX4900R (refurbished)
Hazard: RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
November 8, 2023Medical DevicesFDA
Randox Laboratories Ltd. recalls RX Daytona Plus (without ISE)- IVD Clinical Chemistry Analyser (1) RX4041, (2) RX4041R (refurbished)
Hazard: RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
November 8, 2023Medical DevicesFDA
Randox Laboratories Ltd. recalls RX Daytona Plus (with ISE)-IVD Clinical Chemistry Analyser Catalog Number: RX4040
Hazard: RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
September 27, 2023Medical DevicesFDA
Randox Laboratories Ltd. recalls Liquid Assayed Chemistry Control Premium Plus Level 3, Catalog Number LAE4215
Hazard: there has been a transcription error in the Instructions for Use (IFU) for the Liquid Assayed Chemistry Control Premium Plus Level 3. For Caeruloplasmin, the one and two standard deviation values printed on the sheet are incorrect, which may lead to misreporting patient results. Target and range values are correct, and risk is mitigated; most clinical labs will use the target and range values in clinical practice.
August 30, 2023Medical DevicesFDA
Randox Laboratories Ltd. recalls Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C
Hazard: As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The calibrators lots 2112IT-2116IT, packed into batches 627222, 627224, 634886 and 634887 have been reassigned as part of the restandardization. Following this restandardization, Ferritin results for Quality Control material and patient samples recovered erroneously higher than the targeted calibrator values by approximately +10% across the assay range following.
August 30, 2023Medical DevicesFDA
Randox Laboratories Ltd. recalls Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, C
Hazard: (1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in the Instructions For Use (IFU). The target and ranges for Creatinine for the Roche Creatinine Plus method have been listed incorrectly. These failures cause a delay in patient results for the above mentioned analytes.
August 16, 2023Medical DevicesFDA
Randox Laboratories Ltd. recalls Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena
Hazard: Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after the Direct LDL-cholesterol assay. This may lead to a delay in reporting results or in falsely decreased urea test results.
July 5, 2023Medical DevicesFDA
Randox Laboratories Ltd. recalls Immunoassay Premium Control Tri Level-In vitro diagnostic use, as assayed quality control material to monitor the accura
Hazard: Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has been reassigned on the Roche Cobas e801 due to High recovery outside range
July 5, 2023Medical DevicesFDA
Randox Laboratories Ltd. recalls Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and rep
Hazard: Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has been reassigned on the Roche Cobas e801 due to High recovery outside range
April 5, 2023Medical DevicesFDA
Randox Laboratories recalls CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Cat
Hazard: The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range
January 11, 2023Medical DevicesFDA
Randox Laboratories Ltd. recalls Liquid Assayed Specific Protein Control Level 3- In vitro diagnostic use, in the quality control of serum on clinical ch
Hazard: Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range
January 11, 2023Medical DevicesFDA
Randox Laboratories Ltd. recalls Liquid Assayed Specific Protein Control Level 2- In vitro diagnostic use, in the quality control of serum on clinical c
Hazard: Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range
January 11, 2023Medical DevicesFDA
Randox Laboratories Ltd. recalls Liquid Assayed Specific Protein- In vitro diagnostic use, in the quality control of serum on clinical chemistry and immu
Hazard: Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range
Randox Laboratories Recalls | RecallWatchUSA