Randox Laboratories Ltd. recalls RX Daytona Plus (without ISE)- IVD Clinical Chemistry Analyser (1) RX4041, (2) RX4041R (refurbished)

Recalled November 8, 2023 · FDA recall Z-0199-2024 · Randox Laboratories

Hazard
RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly. · FDA Class II
Units
6 units
Sold at
US Nationwide distribution including Puerto Rico

Description

RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.

Product
RX Daytona Plus (without ISE)- IVD Clinical Chemistry Analyser (1) RX4041, (2) RX4041R (refurbished)
Manufacturer(s)
Randox Laboratories Ltd.
Made in
Ireland