Randox Laboratories Ltd. recalls Calibration Serum Level 3 CAL2351

Recalled January 24, 2024 · FDA recall Z-0752-2024 · Randox Laboratories

Hazard
CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%. · FDA Class II
Units
269 kits
Sold at
US Nationwide and Puerto Rico.

Description

CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.

Product
Calibration Serum Level 3 CAL2351
Manufacturer(s)
Randox Laboratories Ltd.
Made in
Ireland

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov